MATERIALS AND METHODS: The study was performed on Varian Clinac iX, Elekta Synergy S, Siemens Oncor, and Tomotherapy. A CT phantom (Catphan-504, Phantom Laboratory, Salem, NY) was scanned for measurements of image quality including image noise, uniformity, density accuracy, spatial resolution, contrast linearity, and contrast resolution. The measurement results were analysed using in-house image analysis software. Reproducibility, position correction, and geometric accuracy were also evaluated with markers in a smaller alignment phantom. The performance evaluation compared volumetric image properties from these four systems with those from a conventional diagnostic CT (CCT).
RESULTS: It was shown that the linearity of the two kV CBCT was fairly consistent with CCT. The Elekta CBCT with half-circle 27-cm FOV had higher CT numbers than the other three systems. The image noises of the Elekta kV CBCT, Siemens MV CBCT, and Tomotherapy fan-beam CT (FBCT) are about 2-4 times higher than that of the Varian CBCT. The spatial resolutions of two kV CBCTs and two MV CBCTs were 8-11 lp/cm and 3-5 lp/cm, respectively.
CONCLUSION: Elekta CBCT provided a faster image reconstruction and low dose per scan for half-circle scanning. Varian CBCT had relatively lower image noise. Tomotherapy FBCT had the best uniformity.
METHODS: In this international, prospective cohort study of 15,065 patients aged 45 yr or older who underwent in-patient noncardiac surgery, troponin T was measured during the first 3 postoperative days. Patients with a troponin T level of 0.04 ng/ml or greater (elevated "abnormal" laboratory threshold) were assessed for ischemic features (i.e., ischemic symptoms and electrocardiography findings). Patients adjudicated as having a nonischemic troponin elevation (e.g., sepsis) were excluded. To establish diagnostic criteria for MINS, the authors used Cox regression analyses in which the dependent variable was 30-day mortality (260 deaths) and independent variables included preoperative variables, perioperative complications, and potential MINS diagnostic criteria.
RESULTS: An elevated troponin after noncardiac surgery, irrespective of the presence of an ischemic feature, independently predicted 30-day mortality. Therefore, the authors' diagnostic criterion for MINS was a peak troponin T level of 0.03 ng/ml or greater judged due to myocardial ischemia. MINS was an independent predictor of 30-day mortality (adjusted hazard ratio, 3.87; 95% CI, 2.96-5.08) and had the highest population-attributable risk (34.0%, 95% CI, 26.6-41.5) of the perioperative complications. Twelve hundred patients (8.0%) suffered MINS, and 58.2% of these patients would not have fulfilled the universal definition of myocardial infarction. Only 15.8% of patients with MINS experienced an ischemic symptom.
CONCLUSION: Among adults undergoing noncardiac surgery, MINS is common and associated with substantial mortality.