OBJECTIVE: We have conducted a multicenter clinical skin testing study to document the safety and diagnostic sensitivity and specificity of a candidate Hevea brasiliensis nonammoniated latex (NAL) extract. These data are intended to support the licensing of this reagent for the diagnosis of latex allergy in high-risk populations.
METHODS: Three hundred twenty-four subjects (304 adults and 20 children) were classified by their clinical history as having latex allergy (LA group, 124 adults and 10 children) or having no latex allergy (NLA group, 180 adults and 10 children). All subjects provided blood samples and then received sequential puncture skin tests (PSTs) at 1, 100, or 1000 microg/mL protein with a bifurcated needle and NAL (Greer Laboratories) from Malaysian Hevea brasiliensis (clone 600) sap. A 2-stage glove provocation test was used to clarify latex allergy status of individuals with positive history/negative PST result and negative history/positive PST result mismatches.
RESULTS: Twenty-four subjects (15%) originally designated as having LA on the basis of their initial clinical history were reclassified to the NLA group on the basis of a negative glove provocation test result. Of the 134 subjects with LA, 54 (40%) were highly sensitive to latex, with a positive PST result at 1 microg/mL NAL. The Greer NAL reagent produced a positive PST rate (sensitivity) of 95% and 99% in subjects with LA at 100 microg/mL and 1 mg/mL, respectively. The negative PST rate (specificity) in 190 subjects with a negative history with the NAL extract at 100 microg/mL and 1 mg/mL, was 100% and 96%, respectively. Immediately after the PST, mild systemic reactions (mainly pruritus) were recorded in 16.1 % of the adults in the LA group and 4.4% of the adults in the NLA group. No reactions required treatment with epinephrine. Only mild delayed reactions were observed in 9.6% (LA group) and 2.8% (NLA group) of subjects 24 to 48 hours after PST. Mean wheal and erythema diameters measured in the 10 children in the LA group with spina bifida at 100 microg/mL and 1 mg/mL were similar to those observed in the adults in the LA group, suggesting that children are not at increased risk for systemic reactions compared with adults.
CONCLUSIONS: A suggestive clinical history is necessary but not sufficient for a definitive diagnosis of IgE-dependent latex allergy. These data support the safety and diagnostic efficacy of the Greer NAL, skin test reagent at 100 micro/mL and 1 mg/mL for confirmatory PSTs.
OBJECTIVES: We evaluated the relative sensitivity for HPV detection of self-collection compared with practitioner-collected cervical specimens in the context of the Australian National Cervical Screening Program (NCSP).
STUDY DESIGN: 303 women aged ≥18 years attending a single tertiary referral centre took their own sample using a flocked-swab, and then had a practitioner-collected sample taken at colposcopy. All samples were tested at a single laboratory on the six PCR-based HPV assays which can be utilised in the NCSP; Roche cobas 4800 and cobas, Abbott RealTime, BD Onclarity, Cepheid Xpert, and Seegene Anyplex.
RESULTS: HPV16/18 results had high observed agreement between self- and practitioner-collected samples on all assays (range: 0.94-0.99), with good agreement for non-HPV16/18 oncogenic HPV types (range: 0.64-0.73).
CONCLUSIONS: Self-collection for HPV-based cervical screening shows good concordance and relative sensitivity when compared to practitionercollected samples across assays in the NCSP.
METHODS: This evaluation was performed in seven centers across China, South Korea, and Malaysia. Imprecision (repeatability and reproducibility), assay accuracy, and lot-to-lot reagent variability were tested. The Elecsys ECLIAs were compared with commercially available immunoassays (Architect, Dimension, and Viva-E systems) using whole blood samples from patients with various transplant types (kidney, liver, heart, and bone marrow).
RESULTS: Coefficients of variation for repeatability and reproducibility were ≤5.4% and ≤12.4%, respectively, for the tacrolimus ECLIA, and ≤5.1% and ≤7.3%, respectively, for the cyclosporine ECLIA. Method comparisons of the tacrolimus ECLIA with Architect, Dimension, and Viva-E systems yielded slope values of 1.01, 1.14, and 0.897, respectively. The cyclosporine ECLIA showed even closer agreements with the Architect, Dimension, and Viva-E systems (slope values of 1.04, 1.04, and 1.09, respectively). No major differences were observed among the different transplant types.
CONCLUSIONS: The tacrolimus and cyclosporine ECLIAs demonstrated excellent precision and close agreement with other immunoassays tested. These results show that both assays are suitable for ISD monitoring in an APAC population across a range of different transplant types.