METHODS: Liquid based cervico-vaginal cytology specimens with squamous abnormalities and corresponding histology from 97 women with subsequent colposcopy and biopsy were included. The specimens were then subjected to the dual stain and Roche Cobas 4800 multiplex real time PCR HPV DNA testing. The sensitivity and specificity of the dual stain and HPV testing were calculated using CIN 2+ on histology as a reference standard.
RESULTS: The sensitivity and specificity of the dual stain in detecting histology proven CIN 2+ was 93.7% and 76.5% while HPV testing was 85.7% and 14.7% respectively. Of the 44 women with ASCUS or LSIL on cytology, the dual stain also reduced the number of unnecessary colposcopy referrals from 27 to 7 when used as a triage marker compared to HPV testing.
CONCLUSION: p16/Ki-67 dual stain was more sensitive and specific than HPV testing in determining the presence of CIN 2+ on histology. It could triage low grade cervico-vaginal specimens more effectively and potentially help women avoid unnecessary colposcopies. Future studies are needed to further evaluate its role in cervical cancer screening programmes.
METHODS: The halogen lamp was replaced by a low-cost, blue light-emitting diode (LED) lamp. Using a reformulated AO solution, the staining protocol was revised to make use of a concentration gradient instead of uniform staining. To evaluate this new AO diagnostic system, a pilot field study was conducted in the Lake Victoria basin in Kenya.
RESULTS: Without staining failure, malaria infection status of about 100 samples was determined on-site per one microscopist per day, using the improved AO diagnostic system. The improved AO diagnosis had both higher overall sensitivity (46.1 vs 38.9%: p = 0.08) and specificity (99.0 vs 96.3%) than the Giemsa method (N = 1018), using PCR diagnosis as the standard.
CONCLUSIONS: Consistent AO staining of thin blood films and rapid evaluation of malaria parasitaemia with the revised protocol produced superior results relative to the Giemsa method. This AO diagnostic system can be set up easily at low cost using an ordinary light microscope. It may supplement rapid diagnostic tests currently used in clinical settings in malaria-endemic countries, and may be considered as an inexpensive tool for case surveillance in malaria-eliminating countries.