METHODS: A total of 401 whole blood samples with a fresh HbA1c measurement were randomly selected from The Malaysian Cohort's (TMC) biobank. The HbA1c measurements of fresh and frozen (stored for 7-8 years) samples were assayed using different high-performance liquid chromatography (HPLC) systems. The HbA1c values of the fresh samples were then calculated and corrected according to the later system. The reproducibility of HbA1c measurements between calculated-fresh and frozen samples was assessed using a Passing-Bablok linear regression model. The Bland-Altman plot was then used to evaluate the concordance of HbA1c values.
RESULTS: The different HPLC systems highly correlated (r = 0.99) and agreed (ICC = 0.96) with each other. Furthermore, the HbA1c measurements for frozen samples strongly correlate with the corrected HbA1c values of the fresh samples (r = 0.875) with a mean difference of -0.02 (SD: -0.38 to 0.38). Although the mean difference is small, discrepancies were observed within the diabetic and non-diabetic samples.
CONCLUSION: These data demonstrate that the HbA1c measurements between fresh and frozen samples are highly correlated and reproducible.
OBJECTIVE: A sensitive method for the estimation of CRM in plasma, as well as fecal matter-based solid self-nano emulsifying drug delivery system (S-SNEDDS), has been reported for the first time.
METHODS: A bioanalytical method was optimized using Box-Behnken Design having 13 runs and 3 responses. The optimized method was developed using methanol and water (70:30 v/v) with a flow rate of 1 mL/min. Quercetin was used as an internal standard. A specificity test was also performed for the developed CRM solid self-nano emulsifying drug delivery system.
RESULTS: The retention time of CRM was found to be 14.18 minutes. The developed method was validated and found to be linear in the range of 50-250 ng/mL with an R2 of 0.999. Accuracy studies indicated that CRM had a percentage recovery of less than 105% and more than 95%, respectively. Precision studies were carried out for inter, intraday, and inter-analyst precision, and the %RSD was found to be less than 2%. The limit of detection (LOD) and limit of quantification (LOQ) were found to be 3.37 ng/mL and 10.23 ng/mL, respectively. Stability studies for shortterm, long term and freeze-thaw cycles showed a %RSD of less than 2%, indicating the stability of CRM in the plasma matrix. Moreover, the blank fecal microbiota extract slurry did not show any peak at the retention time of CRM in a CRM-loaded solid nanoemulsifying drug delivery system containing fecal microbiota extract indicating its specificity.
CONCLUSION: Hence, the developed method can have clinical implications as it helps estimate CRM in blood samples and also provides a simple and sensitive method for the estimation of plant-based flavonoids along with fecal microbiota extract formulations.