Displaying all 2 publications

Abstract:
Sort:
  1. Eshraghi A, Abu Osman NA, Gholizadeh H, Ali S, Sævarsson SK, Wan Abas WA
    Clin Biomech (Bristol, Avon), 2013 Jan;28(1):55-60.
    PMID: 23157843 DOI: 10.1016/j.clinbiomech.2012.10.002
    Different suspension systems that are used within prosthetic devices may alter the distribution of pressure inside the prosthetic socket in lower limb amputees. This study aimed to compare the interface pressure of a new magnetic suspension system with the pin/lock and Seal-In suspension systems.
  2. Gholizadeh H, Abu Osman NA, Eshraghi A, Ali S, Sævarsson SK, Wan Abas WA, et al.
    J Rehabil Res Dev, 2012;49(9):1321-30.
    PMID: 23408214
    Poor suspension increases slippage of the residual limb inside the socket during ambulation. The main purpose of this article is to evaluate the pistoning at the prosthetic liner-socket interface during gait and assess patients' satisfaction with two different liners. Two prostheses with seal-in and locking liners were fabricated for each of the 10 subjects with transtibial amputation. The Vicon motion system was used to measure the pistoning during gait. The subjects were also asked to complete a Prosthesis Evaluation Questionnaire. The results revealed higher pistoning inside the socket during gait with the locking liner than with the seal-in liner (p < 0.05). The overall satisfaction with the locking liner was higher (p < 0.05) because of the relative ease with which the patients could don and doff the device. As such, pistoning may not be the main factor that determines patients' overall satisfaction with the prosthesis and other factors may also contribute to comfort and satisfaction with prostheses. The article also verifies the feasibility of the Vicon motion system for measuring pistoning during gait.
Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links