  1. Givehchi S, Safari MJ, Tan SK, Md Shah MNB, Sani FBM, Azman RR, et al.
    Phys Med, 2018 Jan;45:198-204.
    PMID: 29373248 DOI: 10.1016/j.ejmp.2017.09.137
    PURPOSE: Accurate determination of the bifurcation angle and correlation with plaque buildup may lead to the prediction of coronary artery disease (CAD). This work evaluates two techniques to measure bifurcation angles in 3D space using coronary computed tomography angiography (CCTA).

    MATERIALS AND METHODS: Nine phantoms were fabricated with different bifurcation angles ranging from 55.3° to 134.5°. General X-ray and CCTA were employed to acquire 2D and 3D images of the bifurcation phantoms, respectively. Multiplanar reformation (MPR) and volume rendering technique (VRT) were used to measure the bifurcation angle between the left anterior descending (LAD) and left circumflex arteries (LCx). The measured angles were compared with the true values to determine the accuracy of each measurement technique. Inter-observer variability was evaluated. The two techniques were further applied on 50 clinical CCTA cases to verify its clinical value.

    RESULTS: In the phantom setting, the mean absolute differences calculated between the true and measured angles by MPR and VRT were 2.4°±2.2° and 3.8°±2.9°, respectively. Strong correlation was found between the true and measured bifurcation angles. Furthermore, no significant differences were found between the bifurcation angles measured using either technique. In clinical settings, large difference of 12.0°±10.6° was found between the two techniques.

    CONCLUSION: In the phantom setting, both techniques demonstrated a significant correlation to the true bifurcation angle. Despite the lack of agreement of the two techniques in the clinical context, our findings in phantoms suggest that MPR should be preferred to VRT for the measurement of coronary bifurcation angle by CCTA.

    Matched MeSH terms: Coronary Angiography/instrumentation
  2. Sabarudin A, Md Yusof AK, Tay MF, Ng KH, Sun Z
    Radiat Prot Dosimetry, 2013;153(4):441-7.
    PMID: 22807493 DOI: 10.1093/rpd/ncs127
    This study was conducted to investigate the effectiveness of dose-saving protocols in dual-source computed tomography (CT) coronary angiography compared with invasive coronary angiography (ICA). On 50 patients who underwent coronary CT angiography was performed dual-source CT (DSCT) and compared with ICA procedures. Entrance skin dose (ESD), which was measured at the thyroid gland, and effective dose (E) were assessed for both imaging modalities. The mean ESD measured at the thyroid gland was the highest at 120 kVp, followed by the 100 kVp DSCT and the ICA protocols with 4.0±1.8, 2.7±1.0 and 1.1±1.2 mGy, respectively. The mean E was estimated to be 10.3±2.1, 6.2±2.3 and 5.3±3.4 mSv corresponding to the 120-kVp, 100-kVp DSCT and ICA protocols, respectively. The application of 100 kVp in DSCT coronary angiography is feasible only in patients with a low body mass index of <25 kg m(-2), which leads to a significant dose reduction with the radiation dose being equivalent to that of ICA.
    Matched MeSH terms: Coronary Angiography/instrumentation*