Displaying 1 publication

Abstract:
Sort:
  1. Rassiah P, Ng KH, DeWerd LA, Kunugi K
    Australas Phys Eng Sci Med, 2004 Mar;27(1):25-9.
    PMID: 15156705
    A thermoluminescent dosimetry (TLD) postal dose inter-comparison was carried out amongst radiotherapy centres in Malaysia. The aim of this TLD inter-comparison was to compare the uniformity involved in the measurement of absorbed dose among the participating centres. A set of 5 TLD chips placed within acrylic trays were mailed to all participating centres for irradiation to an absorbed dose to water of 2 Gy. Measurements were made for 6 MV and 60Co photon beams. Results show an agreement of +/- 5% for all but three radiotherapy centres. The ratios of the TLD readings to that of the reference centre are comparable with other national/regional dose inter-comparisons. The importance of a proper ongoing quality assurance program is essential in maintaining the consistency and uniformity of doses delivered.
    Matched MeSH terms: Radiotherapy Dosage/standards*