Displaying 1 publication

Abstract:
Sort:
  1. Gan SH, Ismail R
    J Chromatogr B Biomed Sci Appl, 2001 Aug 15;759(2):325-35.
    PMID: 11499486
    An HPLC system using solid-phase extraction and HPLC with UV detection has been validated in order to determine tramadol and o-desmethyltramadol (M1) concentrations in human plasma. The method developed was selective and linear for concentrations ranging from 50 to 3,500 ng/ml (tramadol) and 50 to 500 ng/ml (M1) with mean recoveries of 94.36 +/- 12.53% and 93.52 +/- 7.88%, respectively. Limit of quantitation (LOQ) was 50 ng/ml. For tramadol, the intra-day accuracy ranged from 95.48 to 114.64% and the inter-day accuracy, 97.21 to 103.24%. Good precision (0.51 and 18.32% for intra- and inter-day, respectively) was obtained at LOQ. The system has been applied to determine tramadol concentrations in human plasma samples for a pharmacokinetic study.
    Matched MeSH terms: Tramadol/analogs & derivatives*
Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links