METHODS: We present 2 cases in which we encountered premature intravascular detachment of the microcatheter tip and coil migration while treating a dural arteriovenous fistula and aneurysm, respectively. We used a stentriever to remove the detached microcatheter tip and suction using the reperfusion catheter to remove the migrated coil, both techniques that have not been reported in the literature thus far.
RESULTS: Detached microcatheter tip and migrated coil were successfully retrieved using a stentriever and aspiration catheter.
CONCLUSIONS: These novel techniques could potentially reduce mortality and morbidity associated with neurointervention.