METHODS: A total of 83 normal subjects each underwent two visual field examinations with SITA and SPARK on two separate occasions on a randomly selected eye. The eye examined and the order of strategy examined first was randomised but remained constant over the two perimetry visits.
RESULTS: Visual field examination with SPARK Precision was on average 33% faster than SITA Standard. A positive correlation between group mean sensitivities of SITA Standard and SPARK Precision (rho = 0.713, p
METHODS: A total of 39 treated glaucoma patients (34 primary open angle and 5 primary angle closure glaucoma) and 31 cataract patients without glaucoma were tested in succession with the Oculus Twinfield 2 (Oculus Optikgeräte GmbH, Wetzlar, Germany) using the SPARK Precision algorithm and with the HFA 3 (Carl Zeiss Meditec, Dublin, CA) using the 30-2 SITA Standard algorithm.
RESULTS: SPARK Precision required around half the testing time of SITA Standard. There was a good correlation between the MS of the two threshold algorithms but MD and PSD were significantly less severe with SPARK Precision in both glaucoma (focal field loss) and cataract (diffuse field loss) groups (p