  1. Veiskarami A, Malekie S, Kashian S, Tajudin SM
    Rep Pract Oncol Radiother, 2024;29(4):413-425.
    PMID: 39895958 DOI: 10.5603/rpor.101397
    BACKGROUND: Polymer-carbon nanostructures have previously been introduced for dosimetry of gamma rays with potential application in radiotherapy. In this research work, bismuth oxide (Bi2O3) nanoparticles were added into the amorphous polycarbonate (PC) matrix to enhance the probability of the photoelectric effect and dosimetry response in parallel.

    MATERIALS AND METHODS: PC/Bi2O3 nanocomposites at concentrations of 0, 5, 20, 40, and 50 Bi2O3 wt% were fabricated via a solution method. Afterward, the samples were irradiated by gamma rays of cobalt-60 (60Co) related to Picker V-9, and Therarton-780 machines at 30-254 mGy/min. Dosimetric characteristics were carried out including linearity, angular dependency, energy, bias-polarity, field size, and repeatability.

    RESULTS: Field emission scanning electron microscopy (FESEM) and transmission electron microscopy (TEM) analyses exhibited an appropriate dispersion state. The dosimeter response was linear at 30-254 mGy/min for the all samples. The 50 wt% sample exhibited the highest sensitivity at 4.61 nC/mGy. A maximum angular variation of approximately 15% was recorded in normal beam incidence. The energy dependence at two energies of 662 and 1250 keV was obtained as 0.7%. Bias-polarity for the 40, and 50 wt% samples at 400 V were measured as 15.9% and 9.0%, respectively. The dosimetry response was significantly dependent on the radiation field size. Also, the repeatability of the dosimeter response was measured as 0.4%.

    CONCLUSIONS: Considering the dosimetry characteristics of PC-Bi2O3 nanocomposites, and appropriate correction factors, this material can be used as a real-time dosimeter for the photon fields at therapy level.

Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links