  1. Osman ND, Abdulkadir MK, Shuaib IL, Nasirudin RA
    Radiography (Lond), 2024 Jan;30(1):237-244.
    PMID: 38035439 DOI: 10.1016/j.radi.2023.11.012
    INTRODUCTION: The adoption of size-specific dose estimate (SSDE) in clinical practice is still limited owing to the tedious and complex manual measurement of individual patient size for the clinical calculation of SSDE. Thus, the automation of SSDE is imperative. This study aims to evaluate a predictive equation for the automated calculation of SSDE.

    METHODS: A user-friendly software was developed to accurately predict the individual size-specific dose estimation of paediatric patients undergoing computed tomography (CT) scans of the head, thorax, and abdomen. The software includes a calculation equation developed based on a novel SSDE prediction equation that used a population's pre-determined percentage difference between volume-weighted computed tomography dose index (CTDIvol) and SSDE with age. American Association of Physicists in Medicine (AAPM RPT 204) method (manual) and segmentation-based SSDE calculators (indoseCT and XXautocalc) were used to assess the proposed software predictions comparatively.

    RESULTS: The results of this study show that the automated equation-based calculation of SSDE and the manual and segmentation-based calculation of SSDE are in good agreement for patients. The differences between the automated equation-based calculation of SSDE and the manual and segmentation-based calculation are less than 3%.

    CONCLUSION: This study validated an accurate SSDE calculator that allows users to enter key input values and calculate SSDE.

    IMPLICATION FOR PRACTICE: The automated equation-based SSDE software (PESSD) seems a promising tool for estimating individualised CT doses during CT scans.

  2. Abdulkadir MK, Osman ND, Achuthan A, Nasirudin RA, Ahmad MZ, Zain NHM, et al.
    J Med Phys, 2024;49(3):456-463.
    PMID: 39526162 DOI: 10.4103/jmp.jmp_26_24
    BACKGROUND AND PURPOSE: Size-specific dose estimates (SSDE) have been introduced into computed tomography (CT) dosimetry to tailor patients' unique sizes to facilitate accurate CT radiation dose quantification and optimization. The purpose of this study was to develop and validate an automated algorithm for the determination of patient size (effective diameter) and SSDE.

    MATERIALS AND METHODS: A MATLAB platform was used to develop software of algorithms based on image segmentation techniques to automate the calculation of patient size and SSDE. The algorithm was used to automatically estimate the individual size and SSDE of four CT dose index phantoms and 80 CT images of pediatric patients comprising head, thorax, and abdomen scans. For validation, the American Association of Physicists in Medicine (AAPM) manual methods were used to determine the patient's size and SSDE for the same subjects. The accuracy of the proposed algorithm in size and SSDE calculation was evaluated for agreement with the AAPM's estimations (manual) using Bland-Altman's agreement and Pearson's correlation coefficient. The normalized error, system bias, and limits of agreement (LOA) between methods were derived.

    RESULTS: The results demonstrated good agreement and accuracy between the automated and AAPM's patient size estimations with an error rate of 1.9% and 0.27% on the patient and phantoms study, respectively. A 1% percentage difference was found between the automated and manual (AAPM) SSDE estimates. A strong degree of correlation was seen with a narrow LOA between methods for clinical study (r > 0.9771) and phantom study (r > 0.9999).

    CONCLUSION: The proposed automated algorithm provides an accurate estimation of patient size and SSDE with negligible error after validation.