  1. Moey, Soo-Foon, Nur Farah Hani Muhd Jaafar, Nursyahirah Saidin
    MyJurnal
    Introduction: Various medium and high tube potentials were utilized to conduct chest x-rays. There
    are advantages and disadvantages with regards to image quality and radiation dose when using
    medium and high kilovoltage (kVp) technique. However, radiographers have misconstrued
    understanding pertaining to the choice of tube potential as well as grid usage when performing chest radiography. Methods: The experimental study was conducted using the PBU-50 phantom by exposing it with medium kVp utilizing grid and non-grid as well as high kVp with grid. All images obtained were evaluated using the modified evaluation criteria for PA chest established by the Commission of European Communities, 1996 whilst the dose area product (DAP) was determined using the Dose Area Product (DAP) meter. The value obtained from the DAP meter was converted to entrance surface dose (ESD) usingCALDOSE_X5.0 software and mathematical formula. Results: The Wilcoxon Signed-Rank Test indicated a significant difference in ESD when using medium and high kVp; Z= -2.666, p