  1. Kumar Sinasamy T, Nazaruddin Wan Hassan WM, Hardy Mohamad Zaini R, Seevaunnamtum P, Ab Mukmin L
    Anaesthesiol Intensive Ther, 2020;52(5):383-388.
    PMID: 33327696 DOI: 10.5114/ait.2020.101407
    INTRODUCTION: The Baska mask and i-gel are two new types of second-generation supraglottic airway devices. The aim of this study was to compare these two devices in terms of quality of insertion, quality of ventilation and post-insertion complications.

    MATERIAL AND METHODS: A total of 80 adult patients who were scheduled for elective surgery under general anaesthesia were randomised to two groups: Group BM: Baska mask (n = 40) and Group IG: i-gel (n = 40). The assessment focused on ease of insertion, number of attempts, insertion time, number of corrective manoeuvres, oropharyngeal leak pressure, tidal volume, peak airway pressure (PAP) and post-insertion complications.

    RESULTS: Group IG showed a significantly shorter median insertion time (13.3 [interquartile range, IQR 7.8] vs. 17.0 [IQR 9.6] s; P < 0.001), a higher percentage in the 'very easy' ease of insertion category (62.5% vs. 10.0%; P < 0.001), a higher percentage in the no corrective manoeuvre category (92.5% vs. 72.5%; P = 0.003) and a higher percentage in the no post-operative throat pain category (67.5% vs. 32.5%; P = 0.011) than Group BM. However, Group BM showed a significantly higher generated PAP than Group IG (12.7 [1.8] and 11.5 [2.2] cm H2O, respectively; P = 0.010). There were no significant differences in other parameters.

    CONCLUSIONS: The i-gel was better than the Baska mask in terms of ease of insertion, speed of insertion, fewer corrective manoeuvres and less post-operative throat pain. However, the Baska mask had a better cuff seal, as shown by a higher generated PAP.

  2. Narhari R, Nazaruddin Wan Hassan WM, Mohamad Zaini RH, Che Omar S, Abdullah Nik Mohamad N, Seevaunnamtum P
    Anaesthesiol Intensive Ther, 2020;52(5):377-382.
    PMID: 33327695 DOI: 10.5114/ait.2020.101387
    INTRODUCTION: The choice of endotracheal tube (ETT) is important for successful orotracheal fibreoptic intubation (OFI). The aim of this study was to compare the use of the Parker flex tip (PFT) with the unoflex reinforced (UFR) ETT during OFI.

    MATERIAL AND METHODS: A total of 58 patients who underwent elective surgery under general anaesthesia were randomised to two ETT groups, the PFT group (n = 29) and the UFR group (n = 29), for OFI in simulated difficult intubation patients using a rigid cervical collar. After successful standardised induction and relaxation, OFI and railroading of selected ETT were subsequently performed by a similarly experienced practitioner. Ease of insertion, degree of manipulation, time to successful intubation, post-intubation complications and haemodynamic changes were recorded for both groups.

    RESULTS: he percentage of easy intubation was comparable between both groups with a slightly higher percentage in the UFR group than the PFT group (69.0% vs. 62.0%; P = 0.599). Degree of manipulation was also comparable between the two groups; the percentage of cases in which manipulation was not required was slightly higher in the UFR group than the PFT group (69.0% vs. 62.1%; P = 0.849). Time to successful intubation was also comparable between the groups, although the time was slightly shorter for the UFR group than the PFT group (56.9 s ± 39.7 s vs. 63.9 s ± 36.9 s; P = 0.488). There were also no significant differences in other parameters.

    CONCLUSIONS: The Parker flex tip ETT was comparable to the unoflex reinforced ETT for OFI in simulated difficult airway patients.