  1. Verheye S, Wlodarczak A, Montorsi P, Torzewski J, Bennett J, Haude M, et al.
    Catheter Cardiovasc Interv, 2021 Jul 01;98(1):E1-E8.
    PMID: 32881396 DOI: 10.1002/ccd.29260
    OBJECTIVES: We aimed to assess the safety and performance of the Magmaris sirolimus-eluting bioresorbable magnesium scaffold in a large patient population.

    BACKGROUND: Magmaris has shown good outcomes in small-sized controlled trials, but further data are needed to confirm its usability, safety, and performance.

    METHODS: BIOSOLVE-IV is an international, single arm, multicenter registry including patients with a maximum of two single de novo lesions. Follow-up is scheduled up to 5 years; the primary outcome is target lesion failure (TLF) at 12 months.

    RESULTS: A total of 1,075 patients with 1,121 lesions were enrolled. Mean patient age was 61.3 ± 10.5 years and 19.2% (n = 206) presented with non-ST-elevation myocardial infarction (NSTEMI). Lesions were 3.2 ± 0.3 mm in diameter and 14.9 ± 4.2 mm long; 5.1% (n = 57) were bifurcation lesions. Device success was 97.3% (n = 1,129) and procedure success 98.9% (n = 1,063). The Kaplan-Meier estimate of TLF at 12 months was 4.3% [95% confidence interval, CI: 3.2, 5.7] consisting of 3.9% target lesion revascularizations, 0.2% cardiac death, and 1.1% target-vessel myocardial infarction. Definite/probable scaffold thrombosis occurred in five patients (0.5% [95% CI: 0.2, 1.1]), thereof four after early discontinuation of antiplatelet/anticoagulation therapy.

    CONCLUSION: BIOSOLVE-IV confirms the safety and performance of the Magmaris scaffold in a large population with excellent device and procedure success and a very good safety profile up to 12 months in a low-risk population.