In March 2011, an outbreak of acute respiratory disease was reported at the Kuala Lumpur (Malaysia) Police Training Centre. Approximately 100 trainees were hospitalized and 5 were admitted to the intensive care unit. Three of these 5 trainees died. Human adenovirus type 7 was identified as the etiologic agent.
A blocking test was incorporated into the commercial IDEIA Adenovirus test (DAKO Diagnostics Ltd., Cambridgeshire, UK) to detect false positive results when faecal specimens were tested for adenovirus antigen. Immune rabbit serum raised against pooled adenovirus particles from human faecal specimens, together with the pre-immune serum, was used. Assessment of positive showed that false positives were produced under two different conditions: when results were based on visual determination instead of a cut-off value determined from photometric reading, and when absorbance values were not immediately read at the end of the test. Under the optimum condition for reading and assessment of test results (immediate reading and photometric determination), 11% of 65 adenovirus-positive samples were checked by the blocking ELISA as false positives. The rest of the specimens showed blocking of positive absorbance values by 70 to 98%. ELISA was found to be more sensitive than immune electron microscopy on samples with lower antigen concentration.