Displaying all 2 publications

Abstract:
Sort:
  1. Karim MKA, Sabarudin A, Muhammad NA, Ng KH
    Radiol Phys Technol, 2019 Dec;12(4):374-381.
    PMID: 31468370 DOI: 10.1007/s12194-019-00532-8
    This study aimed to evaluate effective dose and size-specific dose estimate (SSDE) of computed tomography angiography (CTA) examination using an anthropomorphic phantom. We included three CTA examination protocols to evaluate the intra- and extra-cranial arteries, pulmonary artery (CTPA), and abdominal vessels. Patient SSDEs were measured retrospectively to estimate patient dose, relative to the bodyweight of the patient and volume CT dose index (CTDIvol). Our findings revealed that the highest dose was absorbed by the left lobe of the thyroid gland during intra-/extra-cranial CTA and CTPA, that is, 14.11 ± 0.24 mGy and 16.20 ± 3.95 mGy, respectively. However, the highest absorbed dose in abdominal/pelvic CTA was the gonads (8.98 ± 0.30 mGy), while other radiosensitive organs in intra- and extra-cranial CTA, CTPA, and abdominal/pelvic CTA did not demonstrate significant differences between organs/structures with p value 0.88, 0.11, and 0.54, respectively. The estimated effective dose in intra-/extra-cranial CTA was lower in patients (0.80 ± 0.60 mSv) than in the phantom (0.83 mSv), but it was the opposite for CTPA, with the effective dose being higher in patients (7.54 ± 3.09 mSv) than in the phantom (6.68 mSv). Similar to the effective dose, only CTPA SSDEs were significantly higher in men than in women (19.74 ± 4.79 mGy versus 7.9 mGy). Effective dose and SSDE are clinically relevant parameters that can help estimate a more accurate patient dose based on a patient's size.
    Matched MeSH terms: Computed Tomography Angiography/instrumentation*
  2. Givehchi S, Safari MJ, Tan SK, Md Shah MNB, Sani FBM, Azman RR, et al.
    Phys Med, 2018 Jan;45:198-204.
    PMID: 29373248 DOI: 10.1016/j.ejmp.2017.09.137
    PURPOSE: Accurate determination of the bifurcation angle and correlation with plaque buildup may lead to the prediction of coronary artery disease (CAD). This work evaluates two techniques to measure bifurcation angles in 3D space using coronary computed tomography angiography (CCTA).

    MATERIALS AND METHODS: Nine phantoms were fabricated with different bifurcation angles ranging from 55.3° to 134.5°. General X-ray and CCTA were employed to acquire 2D and 3D images of the bifurcation phantoms, respectively. Multiplanar reformation (MPR) and volume rendering technique (VRT) were used to measure the bifurcation angle between the left anterior descending (LAD) and left circumflex arteries (LCx). The measured angles were compared with the true values to determine the accuracy of each measurement technique. Inter-observer variability was evaluated. The two techniques were further applied on 50 clinical CCTA cases to verify its clinical value.

    RESULTS: In the phantom setting, the mean absolute differences calculated between the true and measured angles by MPR and VRT were 2.4°±2.2° and 3.8°±2.9°, respectively. Strong correlation was found between the true and measured bifurcation angles. Furthermore, no significant differences were found between the bifurcation angles measured using either technique. In clinical settings, large difference of 12.0°±10.6° was found between the two techniques.

    CONCLUSION: In the phantom setting, both techniques demonstrated a significant correlation to the true bifurcation angle. Despite the lack of agreement of the two techniques in the clinical context, our findings in phantoms suggest that MPR should be preferred to VRT for the measurement of coronary bifurcation angle by CCTA.

    Matched MeSH terms: Computed Tomography Angiography/instrumentation
Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links