  1. Ahmed T, Rahman NA, Alam MK
    Prog Orthod, 2019 Jul 08;20(1):26.
    PMID: 31281954 DOI: 10.1186/s40510-019-0277-x
    BACKGROUND: To introduce an orthodontic bracket debonding device capable of measuring debonding force clinically by a novel sensor mechanism MATERIALS AND METHOD: A prototype orthodontic debonding device was constructed utilizing a lift-off debonding instrument (LODI) and force-sensitive resistor (FSR). For data interpretation, the force sensor was equipped with a microcontroller and C++ programming software running on a computer. Ninety-nine (99) 0.022-in. conventional metallic brackets were bonded to premolar teeth in vitro by a single clinician applying the same adhesive and bonding technique. For validation, the mean debonding force measured by the prototype debonding device (n = 30) and the universal testing machine (n = 30) was compared. Both intra- and inter-examiner reliability tests were done by holding and operating the device in a standardized manner. Following debonding by the prototype device, the bracket failure pattern was evaluated (n = 30) by adhesive remnant index (ARI) under the stereomicroscope at × 30 magnification. Statistical analysis included independent samples t test for validation and intraclass correlation coefficient (ICC) with a 95% confidence interval for both intra- and inter-examiner reliability.

    RESULTS: Mean orthodontic bracket debonding force measured by the prototype device (9.36 ± 1.65 N) and the universal testing machine (10.43 ± 2.71 N) was not significantly different (p 