    PURPOSE: We sought to present our preliminary experience on the effectiveness and safety of magnetic resonance imaging (MRI)-guided, high-intensity focused ultrasound (HIFU) therapy using a volumetric ablation technique in the treatment of Association of Asian Nations (ASEAN) patients with symptomatic uterine leiomyomas.

    METHODS: This study included 33 women who underwent HIFU treatment. Tissue characteristics of leiomyomas were assessed based on T2- and T1-weighted MRI. The immediate nonperfused volume (NPV) ratio and the treatment effectiveness of MRI-guided HIFU on the basis of the degrees of volume reduction and improvement in transformed symptom severity score (SSS) were assessed.

    RESULTS: The median immediate NPV ratio was 89.8%. Additionally, the median acoustic sonication power and HIFU treatment durations were 150 W and 125 min, respectively. At six-month follow-up, the median leiomyoma volume had decreased from 139 mL at baseline to 84 mL and the median transformed SSS had decreased from 56.2 at baseline to 18.8. No major adverse events were observed.

    CONCLUSION: The preliminary results demonstrated that volumetric MRI-guided HIFU therapy for the treatment of symptomatic leiomyomas in ASEAN patients appears to be clinically acceptable with regard to treatment effectiveness and safety.