Protocols developed for high-energy dosimetry IAEA (Technical Reports Series No. 277, 1997), AAPM (Med. Phys. 10 (1983) 741: Med. Phys. 18 (1991) 73: Med. Phys. 21 (1994) 1251), IPEMB (Phys. Med. Biol. 41 (1996) 2557), and HPA (Phys. Med. Biol. 28 (1983) 1097) have continued to enhance precision in dose measurements and the optimization of radiotherapy procedures. While recent dosimetry protocols, including those due to the IAEA and IPEMB, have made a number of improvements compared with previous protocols, it is further desirable to develop absolute dosimetry methods of dose measurements. Measurements based on careful implementation of procedures contained within the various protocols have been carried out in an effort to determine the extent to which discrepancies exist among the protocols. Dose in water at dmax was measured using cylindrical and parallel-plate ionization chambers for 6 MV photon beams and 5 and 12 MeV electron beams. Results obtained from the use of the AAPM and HPA protocols for 6 MV photon beams were found to be 0.9% larger and 0.1% smaller, respectively, than those measured following the IAEA protocol. Calibration dose measurements for 5 and 12 MeV electron beams in water phantoms were found to agree to within 1%, this being well within recommendations from the ICRU and other sources regarding the accuracy of dose delivery.