Abstract

The Contraves Autolyzer 801 is a 20 parameter cell counter that operates on the principle of impedance. This evaluation study found that the within-batch precision for 5 of its parameters--WBC, RBC, Hb, Hct and Plt--had coefficients of variation below 5%, while the counts for red cell parameters were reproducible for up to 3 days. The analyser exhibited good linearity for all the parameters. When compared with 2 other counters--the Coulter S-Plus and the Coulter M350, some degree of proportional and constant bias due to unsuitable calibration were detected in the WBC, Hct and Plt parameters. Otherwise it showed generally good correlation in all the parameters and also with the manual PCV method. Carry over was negligible in the red cell parameters. The Contraves Autolyzer 801 is easy to operate and readily accepted by operators.