Displaying 1 publication

Abstract:
Sort:
  1. Chua, C.S., Fathilah, A.R., Himratul-Aznita, W.H.
    Ann Dent, 2013;20(1):20-26.
    MyJurnal
    Water delivered through dental unit waterline system
    (DUWS) is often reported contaminated with microbes
    dislodged from biofilm that forms within the tubing
    of a dental chair unit (DCU). ADM: The study aimed
    at evaluating the sanitary level of DCU water from
    a teaching dental clinic. Materials: The presence of
    pathogenic bacteria which include total coliforms, faecal
    coliforms, E. coli, faecal streptococci and P. aeruginosa
    were determined using conventional microbiological
    methods while PCR technique was used to identify other
    microbial contaminants. Result: pH of DCU water was
    found slightly acidic at pH 5.4-5.5 and the temperature
    was 23°C. Pathogenic contaminants were absent but
    the DCU water was highly loaded with Sphingomonas
    rhizogenes (17.9%), Sphingomonas dokdonesis
    (79.5%), Sphingomonas mucosissima (1.1%) and
    Methylobacterium radiotolerans (1.5%). The high load of
    microbes that exceeded 200 cfu/ml was of great concern
    as it failed to meet recommendation set by the American
    Dental Association.
Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links