OBJECTIVE: The purpose of this report was to describe the longer-term (>12 months) envelope effects on infection reduction and complications.
METHODS: All trial patients who underwent CIED replacement, upgrade, revision, or initial cardiac resynchronization therapy - defibrillator implantation received standard-of-care infection prophylaxis and were randomized in a 1:1 ratio to receive the envelope or not. CIED infection incidence and procedure and system-related complications were characterized through all follow-up (36 months) by using Cox proportional hazards regression modeling.
RESULTS: In total, 6800 patients received their intended randomized treatment (3371 envelope; 3429 control; mean follow-up period 21.0 ± 8.3 months). Major CIED-related infections occurred in 32 envelope patients and 51 control patients (Kaplan-Meier [KM] estimate 1.3% vs 1.9%; hazard ratio [HR] 0.64; 95% confidence interval [CI] 0.41-0.99; P = .046). Any CIED-related infection occurred in 57 envelope patients and 84 control patients (KM estimate 2.1% vs 2.8%; HR 0.69; 95% CI 0.49-0.97; P = .030). System- or procedure-related complications occurred in 235 envelope patients and 252 control patients (KM estimate 8.0% vs 8.2%; HR 0.95; 95% CI 0.79-1.13; P < .001 for noninferiority); the most common were lead dislodgment (1.1%), device lead damage (0.5%), and implant site hematoma (0.4%). Implant site pain occurred less frequently in the envelope group (0.1% vs 0.4%; P = .067). There were no (0.0%) reports of allergic reactions to the components of the envelope (mesh, polymer, or antibiotics).
CONCLUSION: The effects of the TYRX envelope on the reduction of the risk of CIED infection are sustained beyond the first year postprocedure, without an increased risk of complications.
OBJECTIVE: The purpose of this study was to report the performance of the Micra AV leadless pacemaker from the worldwide Micra AV post-approval registry (PAR) through 12 months.
METHODS: The Micra AV PAR is a prospective, single-arm, observational registry designed to assess the safety and effectiveness of Micra AV in a real-world setting. For the present interim analysis, major complications and system revisions through 12 months were summarized and compared to a historical cohort of 2667 patients implanted with a transvenous dual-chamber pacing system.
RESULTS: The device was successfully implanted in 796 of 801 patients (99.4%) at 97 centers in 19 countries. Micra AV patients were older (74.1 years vs 71.1 years; P < .0001) with a higher incidence of renal disease (22.3% vs 9.8%; P < .0001) compared with transvenous dual-chamber patients. Through 12 months, the major complication rate was 3.7% in Micra AV patients compared with 8.8% in transvenous dual-chamber patients (hazard ratio 0.42; 95% confidence interval 0.28-0.61; P < .001). The system revision rate was 1.5% in Micra AV patients compared with 5.5% for transvenous dual-chamber patients (hazard ratio 0.25; 95% confidence interval 0.13-0.47; P < .001); this reduction was largely driven by the absence of lead dislodgments requiring revision. The median AV synchrony index was 79.4% (interquartile range 65.2%-86.4%) in patients paced >90%.
CONCLUSION: The Micra AV leadless pacemaker was implanted with a high rate of success in patients with multiple comorbidities, with a significantly lower rate of complications and system revisions through 12 months compared to a historical cohort of patients with transvenous dual-chamber pacemakers.