  1. Crossley GH, Biffi M, Johnson B, Lin A, Gras D, Hussin A, et al.
    Heart Rhythm, 2015 Apr;12(4):751-8.
    PMID: 25533587 DOI: 10.1016/j.hrthm.2014.12.019
    The Medtronic Attain Performa quadripolar leads provide 16 pacing vectors with steroid on every electrode. This includes a short bipolar configuration between the middle 2 electrodes.
  2. Yee R, Gadler F, Hussin A, Bin Omar R, Khaykin Y, Verma A, et al.
    Heart Rhythm, 2014 Jul;11(7):1150-5.
    PMID: 24801899 DOI: 10.1016/j.hrthm.2014.04.020
    Left ventricular (LV) lead implantation for cardiac resynchronization therapy (CRT) is associated with lead dislodgement rates ranging from 3% to 10%, and some implant approaches to prevent dislodgement may contribute to suboptimal CRT response. We report our early human experience with an LV lead with a side helix for active fixation to the coronary vein wall.
  3. Mittal S
    Heart Rhythm, 2008 Jun;5(6 Suppl):S64-7.
    PMID: 18456205 DOI: 10.1016/j.hrthm.2008.03.023
    Catheter ablation has come to play an important therapeutic role in the management of some patients with ventricular arrhythmias. An important advance in catheter ablation of ventricular tachycardia (VT) has been the development of three-dimensional catheter-based mapping systems, which permit identification of the tachycardia circuit and facilitate a strategy for catheter ablation. As a result, patients who suffer from "focal" forms of VT have been identified. This can have implications with respect to underlying arrhythmia mechanism, patient prognosis, and therapeutic strategies. The article reviews some insights learned from catheter ablation of focal forms of VT.
  4. Zhang S, Ching CK, Huang D, Liu YB, Rodriguez-Guerrero DA, Hussin A, et al.
    Heart Rhythm, 2020 03;17(3):468-475.
    PMID: 31561030 DOI: 10.1016/j.hrthm.2019.09.023
    BACKGROUND: Implantable cardioverter-defibrillators (ICDs) are underutilized in Asia, Latin America, Eastern Europe, the Middle East, and Africa. The Improve SCA Study is the largest prospective study to evaluate the benefit of ICD therapy in underrepresented geographies. This analysis reports the primary objective of the study.

    OBJECTIVES: The objectives of this study was to determine whether patients with primary prevention (PP) indications with specific risk factors (1.5PP: syncope, nonsustained ventricular tachycardia, premature ventricular contractions >10/h, and low ventricular ejection fraction <25%) are at a similar risk of life-threatening arrhythmias as patients with secondary prevention (SP) indications and to evaluate all-cause mortality rates in 1.5PP patients with and without devices.

    METHODS: A total of 3889 patients were included in the analysis to evaluate ventricular tachycardia or fibrillation therapy and mortality rates. Patients were stratified as SP (n = 1193) and patients with PP indications. The PP cohort was divided into 1.5PP patients (n = 1913) and those without any 1.5PP criteria (n = 783). The decision to undergo ICD implantation was left to the patient and/or physician. The Cox proportional hazards model was used to compute hazard ratios.

    RESULTS: Patients had predominantly nonischemic cardiomyopathy. The rate of ventricular tachycardia or fibrillation in 1.5PP patients was not equivalent (within 30%) to that in patients with SP indications (hazard ratio 0.47; 95% confidence interval 0.38-0.57) but was higher than that in PP patients without any 1.5PP criteria (hazard ratio 0.67; 95% confidence interval 0.46-0.97) (P = .03). There was a 49% relative risk reduction in all-cause mortality in ICD implanted 1.5PP patients. In addition, the number needed to treat to save 1 life over 3 years was 10.0 in the 1.5PP cohort vs 40.0 in PP patients without any 1.5PP criteria.

    CONCLUSION: These data corroborate the mortality benefit of ICD therapy and support extension to a selected PP population from underrepresented geographies.

  5. Mittal S, Wilkoff BL, Kennergren C, Poole JE, Corey R, Bracke FA, et al.
    Heart Rhythm, 2020 07;17(7):1115-1122.
    PMID: 32087357 DOI: 10.1016/j.hrthm.2020.02.011
    BACKGROUND: The World-wide Randomized Antibiotic Envelope Infection Prevention trial reported a 40% reduction in major cardiac implantable electronic device (CIED) infections within 12 months of the procedure with the use of an antibacterial-eluting envelope (TYRX Absorbable Antibacterial Envelope, Medtronic, Mounds View, MN).

    OBJECTIVE: The purpose of this report was to describe the longer-term (>12 months) envelope effects on infection reduction and complications.

    METHODS: All trial patients who underwent CIED replacement, upgrade, revision, or initial cardiac resynchronization therapy - defibrillator implantation received standard-of-care infection prophylaxis and were randomized in a 1:1 ratio to receive the envelope or not. CIED infection incidence and procedure and system-related complications were characterized through all follow-up (36 months) by using Cox proportional hazards regression modeling.

    RESULTS: In total, 6800 patients received their intended randomized treatment (3371 envelope; 3429 control; mean follow-up period 21.0 ± 8.3 months). Major CIED-related infections occurred in 32 envelope patients and 51 control patients (Kaplan-Meier [KM] estimate 1.3% vs 1.9%; hazard ratio [HR] 0.64; 95% confidence interval [CI] 0.41-0.99; P = .046). Any CIED-related infection occurred in 57 envelope patients and 84 control patients (KM estimate 2.1% vs 2.8%; HR 0.69; 95% CI 0.49-0.97; P = .030). System- or procedure-related complications occurred in 235 envelope patients and 252 control patients (KM estimate 8.0% vs 8.2%; HR 0.95; 95% CI 0.79-1.13; P < .001 for noninferiority); the most common were lead dislodgment (1.1%), device lead damage (0.5%), and implant site hematoma (0.4%). Implant site pain occurred less frequently in the envelope group (0.1% vs 0.4%; P = .067). There were no (0.0%) reports of allergic reactions to the components of the envelope (mesh, polymer, or antibiotics).

    CONCLUSION: The effects of the TYRX envelope on the reduction of the risk of CIED infection are sustained beyond the first year postprocedure, without an increased risk of complications.

  6. Garweg C, Khelae SK, Steinwender C, Chan JYS, Ritter P, Johansen JB, et al.
    Heart Rhythm, 2020 12;17(12):2037-2045.
    PMID: 32717315 DOI: 10.1016/j.hrthm.2020.07.024
    BACKGROUND: The MARVEL (Micra Atrial TRacking Using a Ventricular AccELerometer) 2 study assessed the efficacy of atrioventricular (AV) synchronous pacing with a Micra leadless pacemaker. Average atrioventricular synchrony (AVS) was 89.2%. Previously, low amplitude of the Micra-sensed atrial signal (A4) was observed to be a factor of low AVS.

    OBJECTIVE: The purpose of this study was to identify predictors of A4 amplitude and high AVS.

    METHODS: We analyzed 64 patients enrolled in MARVEL 2 who had visible P waves on electrocardiogram for assessing A4 amplitude and 40 patients with third-degree AV block for assessing AVS at rest. High AVS was defined as >90% correct atrial-triggered ventricular pacing. The association between clinical factors and echocardiographic parameters with A4 amplitude was investigated using a multivariable model with lasso variable selection. Variables associated with A4 amplitude together with premature ventricular contraction burden, sinus rate, and sinus rate variability (standard deviation of successive differences of P-P intervals [SDSD]) were assessed for association with AVS.

    RESULTS: In univariate analysis, low A4 amplitude was inversely related to atrial function assessed by E/A ratio and e'/a' ratio, and was directly related to atrial contraction excursion (ACE) and atrial strain (Ɛa) on echocardiography (all P ≤.05). The multivariable lasso regression model found coronary artery bypass graft history, E/A ratio, ACE, and Ɛa were associated with low A4 amplitude. E/A ratio and SDSD were multivariable predictors of high AVS, with >90% probability if E/A <0.94 and SDSD <5 bpm.

    CONCLUSION: Clinical parameters and echocardiographic markers of atrial function are associated with A4 signal amplitude. High AVS can be predicted by E/A ratio <0.94 and low sinus rate variability at rest.

  7. Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al.
    Heart Rhythm, 2017 05;14(5):685-691.
    PMID: 28111349 DOI: 10.1016/j.hrthm.2017.01.026
    BACKGROUND: Device repositioning during Micra leadless pacemaker implantation may be required to achieve optimal pacing thresholds.

    OBJECTIVE: The purpose of this study was to describe the natural history of acute elevated Micra vs traditional transvenous lead thresholds.

    METHODS: Micra study VVI patients with threshold data (at 0.24 ms) at implant (n = 711) were compared with Capture study patients with de novo transvenous leads at 0.4 ms (n = 538). In both cohorts, high thresholds were defined as >1.0 V and very high as >1.5 V. Change in pacing threshold (0-6 months) with high (1.0 to ≤1.5 V) or very high (>1.5 V) thresholds were compared using the Wilcoxon signed-rank test.

    RESULTS: Of the 711 Micra patients, 83 (11.7%) had an implant threshold of >1.0 V at 0.24 ms. Of the 538 Capture patients, 50 (9.3%) had an implant threshold of >1.0 V at 0.40 ms. There were no significant differences in patient characteristics between those with and without an implant threshold of >1.0 V, with the exception of left ventricular ejection fraction in the Capture cohort (high vs low thresholds, 53% vs 58%; P = .011). Patients with an implant threshold of >1.0 V decreased significantly (P < .001) in both cohorts. Micra patients with high and very high thresholds decreased significantly (P < .01) by 1 month, with 87% and 85% having 6-month thresholds lower than the implant value. However, when the capture threshold at implant was >2 V, only 18.2% had a threshold of ≤1 V at 6 months and 45.5% had a capture threshold of >2 V.

    CONCLUSIONS: Pacing thresholds in most Micra patients with elevated thresholds decrease after implant. Micra device repositioning may not be necessary if the pacing threshold is ≤2 V.

  8. Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JM, et al.
    Heart Rhythm, 2017 05;14(5):702-709.
    PMID: 28192207 DOI: 10.1016/j.hrthm.2017.01.035
    BACKGROUND: Early performance of the Micra transcatheter pacemaker from the global clinical trial reported a 99.2% implant success rate, low and stable pacing capture thresholds, and a low (4.0%) rate of major complications up to 6 months.

    OBJECTIVE: The purpose of this report was to describe the prespecified long-term safety objective of Micra at 12 months and electrical performance through 24 months.

    METHODS: The Micra Transcatheter Pacing Study was a prospective single-arm study designed to assess the safety and efficacy of the Micra VVIR leadless/intracardiac pacemaker. Enrolled patients met class I or II guideline recommendations for de novo ventricular pacing. The long-term safety objective was freedom from a system- or procedure-related major complication at 12 months. A predefined historical control group of 2667 patients with transvenous pacemakers was used to compare major complication rates.

    RESULTS: The long-term safety objective was achieved with a freedom from major complication rate of 96.0% at 12 months (95% confidence interval 94.2%-97.2%; P < .0001 vs performance goal). The risk of major complications for patients with Micra (N = 726) was 48% lower than that for patients with transvenous systems through 12 months postimplant (hazard ratio 0.52; 95% confidence interval 0.35-0.77; P = .001). Across subgroups of age, sex, and comorbidities, Micra reduced the risk of major complications compared to transvenous systems. Electrical performance was excellent through 24 months, with a projected battery longevity of 12.1 years.

    CONCLUSION: Long-term performance of the Micra transcatheter pacemaker remains consistent with previously reported data. Few patients experienced major complications through 12 months of follow-up, and all patient subgroups benefited as compared to transvenous pacemaker historical control group.

  9. Chinitz L, Ritter P, Khelae SK, Iacopino S, Garweg C, Grazia-Bongiorni M, et al.
    Heart Rhythm, 2018 09;15(9):1363-1371.
    PMID: 29758405 DOI: 10.1016/j.hrthm.2018.05.004
    BACKGROUND: Micra is a leadless pacemaker that is implanted in the right ventricle and provides rate response via a 3-axis accelerometer (ACC). Custom software was developed to detect atrial contraction using the ACC enabling atrioventricular (AV) synchronous pacing.

    OBJECTIVE: The purpose of this study was to sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing.

    METHODS: The Micra Accelerometer Sensor Sub-Study (MASS) and MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. The Micra Atrial TRacking Using A Ventricular AccELerometer (MARVEL) study was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. Atrioventricular synchrony (AVS) was measured during 30 minutes of rest and during VVI pacing. AVS was defined as a P wave visible on surface ECG followed by a ventricular event <300 ms.

    RESULTS: A total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range 0-41.4). High-degree AV block was present in 33 patients, whereas 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (95% confidence interval 81.8%-90.9%), 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P

  10. Lloyd M, Reynolds D, Sheldon T, Stromberg K, Hudnall JH, Demmer WM, et al.
    Heart Rhythm, 2017 02;14(2):200-205.
    PMID: 27871854 DOI: 10.1016/j.hrthm.2016.11.016
    BACKGROUND: The Micra transcatheter pacemaker was designed to have similar functionality to conventional transvenous VVIR pacing systems. It provides rate adaptive pacing using a programmable 3-axis accelerometer designed to detect patient activity in the presence of cardiac motion.

    OBJECTIVE: The purpose of this study was to evaluate the system's performance during treadmill tests to maximum exertion in a subset of patients within the Micra Transcatheter Pacing Study.

    METHODS: Patients underwent treadmill testing at 3 or 6 months postimplant with algorithm programming at physician discretion. Normalized sensor rate (SenR) relative to the programmed upper sensor rate was modeled as a function of normalized workload in metabolic equivalents (METS) relative to maximum METS achieved during the test. A normalized METS and SenR were determined at the end of each 1-minute treadmill stage. The proportionality of SenR to workload was evaluated by comparing the slope from this relationship to the prospectively defined tolerance margin (0.65-1.35).

    RESULTS: A total of 69 treadmill tests were attempted by 42 patients at 3 and 6 months postimplant. Thirty tests from 20 patients who completed ≥4 stages with an average slope of 0.86 (90% confidence interval 0.77-0.96) confirmed proportionality to workload. On an individual test basis, 25 of 30 point estimates (83.3%) had a normalized slope within the defined tolerance range (range 0.46-1.08).

    CONCLUSION: Accelerometer-based rate adaptive pacing was proportional to workload, thus confirming rate adaptive pacing commensurate to workload is achievable with an entirely intracardiac pacing system.

  11. Afzal MR, Daoud EG, Cunnane R, Mulpuru SK, Koay A, Hussain A, et al.
    Heart Rhythm, 2018 06;15(6):841-846.
    PMID: 29427820 DOI: 10.1016/j.hrthm.2018.02.008
    BACKGROUND: Experience with retrieval of the Micra transcatheter pacing system (TPS) is limited because of its relatively newer technology. Although abandonment of the TPS at end of life is recommended, certain situations such as endovascular infection or device embolization warrant retrieval.

    OBJECTIVE: The purpose of this study was to report the worldwide experience with successful retrieval of the Micra TPS.

    METHODS: A list of all successful retrievals of the currently available leadless pacemakers (LPs) was obtained from the manufacturer of Micra TPS. Pertinent details of retrieval, such as indication, days postimplantation, equipment used, complications, and postretrieval management, were obtained from the database collected by the manufacturer. Other procedural details were obtained directly from the operators at each participating site.

    RESULTS: Data from the manufacturer consisted of 40 successful retrievals of the Micra TPS. Operators for 29 retrievals (73%) provided the consent and procedural details. Of the 29 retrievals, 11 patients underwent retrieval during the initial procedure (immediate retrieval); the other 18 patients underwent retrieval during a separate procedure (delayed retrieval). Median duration before delayed retrieval was 46 days (range 1-95 days). The most common reason for immediate retrieval was elevated pacing threshold after tether removal. The most common reasons for delayed retrieval included elevated pacing threshold at follow-up, endovascular infection, and need for transvenous device. Mean procedure duration was 63.11 ± 56 minutes. All retrievals involved snaring via a Micra TPS delivery catheter or steerable sheath. No serious complications occurred during the reported retrievals.

    CONCLUSION: Early retrieval of the Micra TPS is feasible and safe.