OBJECTIVE: The purpose of this report was to describe the prespecified long-term safety objective of Micra at 12 months and electrical performance through 24 months.
METHODS: The Micra Transcatheter Pacing Study was a prospective single-arm study designed to assess the safety and efficacy of the Micra VVIR leadless/intracardiac pacemaker. Enrolled patients met class I or II guideline recommendations for de novo ventricular pacing. The long-term safety objective was freedom from a system- or procedure-related major complication at 12 months. A predefined historical control group of 2667 patients with transvenous pacemakers was used to compare major complication rates.
RESULTS: The long-term safety objective was achieved with a freedom from major complication rate of 96.0% at 12 months (95% confidence interval 94.2%-97.2%; P < .0001 vs performance goal). The risk of major complications for patients with Micra (N = 726) was 48% lower than that for patients with transvenous systems through 12 months postimplant (hazard ratio 0.52; 95% confidence interval 0.35-0.77; P = .001). Across subgroups of age, sex, and comorbidities, Micra reduced the risk of major complications compared to transvenous systems. Electrical performance was excellent through 24 months, with a projected battery longevity of 12.1 years.
CONCLUSION: Long-term performance of the Micra transcatheter pacemaker remains consistent with previously reported data. Few patients experienced major complications through 12 months of follow-up, and all patient subgroups benefited as compared to transvenous pacemaker historical control group.
OBJECTIVE: The purpose of this study was to report the performance of the Micra AV leadless pacemaker from the worldwide Micra AV post-approval registry (PAR) through 12 months.
METHODS: The Micra AV PAR is a prospective, single-arm, observational registry designed to assess the safety and effectiveness of Micra AV in a real-world setting. For the present interim analysis, major complications and system revisions through 12 months were summarized and compared to a historical cohort of 2667 patients implanted with a transvenous dual-chamber pacing system.
RESULTS: The device was successfully implanted in 796 of 801 patients (99.4%) at 97 centers in 19 countries. Micra AV patients were older (74.1 years vs 71.1 years; P < .0001) with a higher incidence of renal disease (22.3% vs 9.8%; P < .0001) compared with transvenous dual-chamber patients. Through 12 months, the major complication rate was 3.7% in Micra AV patients compared with 8.8% in transvenous dual-chamber patients (hazard ratio 0.42; 95% confidence interval 0.28-0.61; P < .001). The system revision rate was 1.5% in Micra AV patients compared with 5.5% for transvenous dual-chamber patients (hazard ratio 0.25; 95% confidence interval 0.13-0.47; P < .001); this reduction was largely driven by the absence of lead dislodgments requiring revision. The median AV synchrony index was 79.4% (interquartile range 65.2%-86.4%) in patients paced >90%.
CONCLUSION: The Micra AV leadless pacemaker was implanted with a high rate of success in patients with multiple comorbidities, with a significantly lower rate of complications and system revisions through 12 months compared to a historical cohort of patients with transvenous dual-chamber pacemakers.