  1. Singh OP, Howe TA, Malarvili MB
    J Breath Res, 2018 01 04;12(2):026003.
    PMID: 28928295 DOI: 10.1088/1752-7163/aa8dbd
    The development of a human respiration carbon dioxide (CO2) measurement device to evaluate cardiorespiratory status inside and outside a hospital setting has proven to be a challenging area of research over the few last decades. Hence, we report a real-time, user operable CO2 measurement device using an infrared CO2 sensor (Arduino Mega2560) and a thin film transistor (TFT, 3.5″), incorporated with low pass (cut-off frequency, 10 Hz) and moving average (span, 8) filters. The proposed device measures features such as partial end-tidal carbon dioxide (EtCO2), respiratory rate (RR), inspired carbon dioxide (ICO2), and a newly proposed feature-Hjorth activity-that annotates data with the date and time from a real-time clock, and is stored onto a secure digital (SD) card. Further, it was tested on 22 healthy subjects and the performance (reliability, validity and relationship) of each feature was established using (1) an intraclass correlation coefficient (ICC), (2) standard error measurement (SEM), (3) smallest detectable difference (SDD), (4) Bland-Altman plot, and (5) Pearson's correlation (r). The SEM, SDD, and ICC values for inter- and intra-rater reliability were less than 5% and more than 0.8, respectively. Further, the Bland-Altman plot demonstrates that mean differences ± standard deviations for a set limit were 0.30 ± 0.77 mmHg, -0.34 ± 1.41 mmHg and 0.21 ± 0.64 breath per minute (bpm) for CO2, EtCO2 and RR. The findings revealed that the developed device is highly reliable, providing valid measurements for CO2, EtCO2, ICO2 and RR, and can be used in clinical settings for cardiorespiratory assessment. This research also demonstrates that EtCO2 and RR (r, -0.696) are negatively correlated while EtCO2 and activity (r, 0.846) are positively correlated. Thus, simultaneous measurement of these features may possibly assist physicians in understanding the subject's cardiopulmonary status. In future, the proposed device will be tested with asthmatic patients for use as an early screening tool outside a hospital setting.
  2. Howe TA, Jaalam K, Ahmad R, Sheng CK, Nik Ab Rahman NH
    J Emerg Med, 2011 Dec;41(6):581-9.
    PMID: 19272745 DOI: 10.1016/j.jemermed.2008.10.017
    STUDY OBJECTIVE: To determine if the slope of Phase II and Phase III, and the alpha angle of the expiratory capnographic waveform, as measured via computer-recognizable algorithms, can reflect changes in bronchospasm in acute asthmatic non-intubated patients presenting to the emergency department (ED).
    METHODS: In this prospective study carried out in a university hospital ED, 30 patients with acute asthma were monitored with clinical severity scoring and peak flow measurements, and then had a nasal cannula attached for sidestream sampling of expired carbon dioxide. The capnographic waveform was recorded onto a personal computer card for analysis. The patients were treated according to departmental protocols. After treatment, when they had improved enough for discharge, a second set of results was obtained for capnographic waveform recording. The pre-treatment and post-treatment results were then compared with paired-samples t-test analysis.
    RESULTS: On the capnographic waveform pre- and post-treatment, there was a significant difference in the slope of Phase III (p < 0.001) and alpha angle (p < 0.001), but not in the Phase II slope (p = 0.35). There was significant change in peak flow meter reading, but it was poorly correlated with all the capnographic indices.
    CONCLUSION: The study provides some preliminary data showing that capnographic waveform indices can indicate improvement in airway diameter in acute asthmatics in the ED. Capnographic waveform analysis presents several advantages in that it is effort-independent, and provides continuous monitoring of normal tidal respiration. With further refined studies, it may serve as a new method of monitoring non-intubated asthmatics in the ED.
    Study site: Emergency department, Hospital Universiti Sains Malaysia (HUSM), Kubang Kerian, Kelantan, Malaysia