Displaying 1 publication

Abstract:
Sort:
  1. Wong WR, Krupin O, Sekaran SD, Mahamd Adikan FR, Berini P
    Anal Chem, 2014 Feb 4;86(3):1735-43.
    PMID: 24410440 DOI: 10.1021/ac403539k
    We present a compact, cost-effective, label-free, real-time biosensor based on long-range surface plasmon polariton (LRSPP) gold (Au) waveguides for the detection of dengue-specific immunoglobulin M (IgM) antibody, and we demonstrate detection in actual patient blood plasma samples. Two surface functionalization approaches are proposed and demonstrated: a dengue virus serotype 2 (DENV-2) functionalized surface to capture dengue-specific IgM antibody in blood plasma and the reverse, a blood plasma functionalized surface to capture DENV-2. The results obtained via these two surface functionalization approaches are comparable to, or of greater quality, than those collected by conventional IgM antibody capture enzyme linked immunosorbent assay (MAC-ELISA). Our second functionalization approach was found to minimize nonspecific binding, thus improving the sensitivity and accuracy of the test. We also demonstrate reuse of the biosensors by regenerating the sensing surface down to the virus (or antibody) level or down to the bare Au.
Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links