Displaying all 2 publications

Abstract:
Sort:
  1. Noriah Jamal, Ng, Kwan-Hoong, McLean, Donald
    MyJurnal
    The objective of this study is to compare low-contrast detectability of computed radiography (CR) and screen/film (SF) mammography systems. The Nijimegen contrast detail test object (CDMAM type 3.4) was imaged at 28 kV, in automatic exposure control mode separately. Six medical imaging physicists read each CDMAM phantom image. Contrast detail curves were plotted to compare lowcontrast detectability of CR (soft copy and hard copy) and SF mammography systems. Effect of varying exposure parameters, namely kV, object position inside the breast phantom, and entrance surface exposure (ESE) on the contrast detail curve were also investigated using soft copy CR. The significance of the difference in contrast between CR and SF, and for each exposure parameter, was tested using non-parametric Kruskal-Wallis test. The low- contrast detectability of the CR (soft copy and hard copy) system was found to be not significantly different to that of the SF system (p>0.05, Kruskal-Wallis test). For CR soft copy, no significant relationship (p>0.05, Kruskal-Wallis test) was seen for variation of kV, object position inside the breast phantom and ESE. This indicates that CR is comparable with SF for useful detection and visualization of low-contrast objects such as small lowcontrast areas corresponding to breast pathology.
  2. Jamal N, Ng KH, McLean D, Looi LM, Moosa F
    AJR Am J Roentgenol, 2004 Mar;182(3):713-7.
    PMID: 14975974
    This study was undertaken to estimate mammographic breast glandularity in Malaysian women from radiographic data.
Related Terms
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links