  1. Chinitz L, Ritter P, Khelae SK, Iacopino S, Garweg C, Grazia-Bongiorni M, et al.
    Heart Rhythm, 2018 09;15(9):1363-1371.
    PMID: 29758405 DOI: 10.1016/j.hrthm.2018.05.004
    BACKGROUND: Micra is a leadless pacemaker that is implanted in the right ventricle and provides rate response via a 3-axis accelerometer (ACC). Custom software was developed to detect atrial contraction using the ACC enabling atrioventricular (AV) synchronous pacing.

    OBJECTIVE: The purpose of this study was to sense atrial contractions from the Micra ACC signal and provide AV synchronous pacing.

    METHODS: The Micra Accelerometer Sensor Sub-Study (MASS) and MASS2 early feasibility studies showed intracardiac accelerations related to atrial contraction can be measured via ACC in the Micra leadless pacemaker. The Micra Atrial TRacking Using A Ventricular AccELerometer (MARVEL) study was a prospective multicenter study designed to characterize the closed-loop performance of an AV synchronous algorithm downloaded into previously implanted Micra devices. Atrioventricular synchrony (AVS) was measured during 30 minutes of rest and during VVI pacing. AVS was defined as a P wave visible on surface ECG followed by a ventricular event <300 ms.

    RESULTS: A total of 64 patients completed the MARVEL study procedure at 12 centers in 9 countries. Patients were implanted with a Micra for a median of 6.0 months (range 0-41.4). High-degree AV block was present in 33 patients, whereas 31 had predominantly intrinsic conduction during the study. Average AVS during AV algorithm pacing was 87.0% (95% confidence interval 81.8%-90.9%), 80.0% in high-degree block patients and 94.4% in patients with intrinsic conduction. AVS was significantly greater (P

  2. Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al.
    N Engl J Med, 2016 Feb 11;374(6):533-41.
    PMID: 26551877 DOI: 10.1056/NEJMoa1511643
    BACKGROUND: A leadless intracardiac transcatheter pacing system has been designed to avoid the need for a pacemaker pocket and transvenous lead.
    METHODS: In a prospective multicenter study without controls, a transcatheter pacemaker was implanted in patients who had guideline-based indications for ventricular pacing. The analysis of the primary end points began when 300 patients reached 6 months of follow-up. The primary safety end point was freedom from system-related or procedure-related major complications. The primary efficacy end point was the percentage of patients with low and stable pacing capture thresholds at 6 months (≤2.0 V at a pulse width of 0.24 msec and an increase of ≤1.5 V from the time of implantation). The safety and efficacy end points were evaluated against performance goals (based on historical data) of 83% and 80%, respectively. We also performed a post hoc analysis in which the rates of major complications were compared with those in a control cohort of 2667 patients with transvenous pacemakers from six previously published studies.
    RESULTS: The device was successfully implanted in 719 of 725 patients (99.2%). The Kaplan-Meier estimate of the rate of the primary safety end point was 96.0% (95% confidence interval [CI], 93.9 to 97.3; P<0.001 for the comparison with the safety performance goal of 83%); there were 28 major complications in 25 of 725 patients, and no dislodgements. The rate of the primary efficacy end point was 98.3% (95% CI, 96.1 to 99.5; P<0.001 for the comparison with the efficacy performance goal of 80%) among 292 of 297 patients with paired 6-month data. Although there were 28 major complications in 25 patients, patients with transcatheter pacemakers had significantly fewer major complications than did the control patients (hazard ratio, 0.49; 95% CI, 0.33 to 0.75; P=0.001).
    CONCLUSIONS: In this historical comparison study, the transcatheter pacemaker met the prespecified safety and efficacy goals; it had a safety profile similar to that of a transvenous system while providing low and stable pacing thresholds. (Funded by Medtronic; Micra Transcatheter Pacing Study ClinicalTrials.gov number, NCT02004873.).