  1. Warrener L, Slibinskas R, Chua KB, Nigatu W, Brown KE, Sasnauskas K, et al.
    Bull World Health Organ, 2011 Sep 01;89(9):675-82.
    PMID: 21897488 DOI: 10.2471/BLT.11.088427
    OBJECTIVE: To evaluate the performance of a newly developed point-of-care test (POCT) for the detection of measles-specific IgM antibodies in serum and oral fluid specimens and to assess if measles virus nucleic acid could be recovered from used POCT strips.

    METHODS: The POCT was used to test 170 serum specimens collected through measles surveillance or vaccination programmes in Ethiopia, Malaysia and the Russian Federation: 69 were positive for measles immunoglobulin M (IgM) antibodies, 74 were positive for rubella IgM antibodies and 7 were positive for both. Also tested were 282 oral fluid specimens from the measles, mumps and rubella (MMR) surveillance programme of the United Kingdom of Great Britain and Northern Ireland. The Microimmune measles IgM capture enzyme immunoassay was the gold standard for comparison. A panel of 24 oral fluids was used to investigate if measles virus haemagglutinin (H) and nucleocapsid (N) genes could be amplified by polymerase chain reaction directly from used POCT strips.

    FINDINGS: With serum POCT showed a sensitivity and specificity of 90.8% (69/76) and 93.6% (88/94), respectively; with oral fluids, sensitivity and specificity were 90.0% (63/70) and 96.2% (200/208), respectively. Both H and N genes were reliably detected in POCT strips and the N genes could be sequenced for genotyping. Measles virus genes could be recovered from POCT strips after storage for 5 weeks at 20-25 °C.

    CONCLUSION: The POCT has the sensitivity and specificity required of a field-based test for measles diagnosis. However, its role in global measles control programmes requires further evaluation.

  2. Senin A, Noordin NM, Sani JAM, Mahat D, Donadel M, Scobie HM, et al.
    PLoS One, 2024;19(3):e0298730.
    PMID: 38483868 DOI: 10.1371/journal.pone.0298730
    INTRODUCTION: A lateral flow rapid diagnostic test (RDT) enables detection of measles specific immunoglobulin M (IgM) antibody in serum, capillary blood, and oral fluid with accuracy consistent with enzyme immunoassay (EIA). The objectives of the study were: 1) to assess measles RDT inter-reader agreement between two clinic staff; 2) to assess the sensitivity and specificity of the measles RDT relative to standard surveillance testing in a low transmission setting; 3) to evaluate the knowledge, attitudes, and practices of staff in clinics using the RDT; and 4) to assess the impact of RDT testing on the measles public health response in Malaysia.

    MATERIALS AND METHODS: The clinic-based prospective evaluation included all suspected measles cases captured by routine measles surveillance at 34 purposely selected clinics in 15 health districts in Malaysia between September 2019 and June 2020, following day-long regional trainings on RDT use. Following informed consent, four specimens were collected from each suspected case, including those routinely collected for standard surveillance [serum for EIA and throat swabs for quantitative reverse transcriptase polymerase chain reaction (RT-qPCR)] together with capillary blood and oral fluid tested with RDTs during the study. RDT impact was evaluated by comparing the rapidity of measles public health response between the pre-RDT implementation (December 2018 to August 2019) and RDT implementation periods (September 2019 to June 2020). To assess knowledge, attitudes, and practices of RDT use, staff involved in the public health management of measles at the selected sites were surveyed.

    RESULTS: Among the 436 suspect cases, agreement of direct visual readings of measles RDT devices between two health clinic staff was 99% for capillary blood (k = 0.94) and 97% for oral fluid (k = 0.90) specimens. Of the total, 45 (10%) were positive by measles IgM EIA (n = 44, including five also positive by RT-qPCR) or RT-qPCR only (n = 1), and 38 were positive by RDT (using either capillary blood or oral fluid). Using measles IgM EIA or RT-qPCR as reference, RDT sensitivity using capillary blood was 43% (95% CI: 30%-58%) and specificity was 98% (95% CI: 96%-99%); using oral fluid, sensitivity (26%, 95% CI: 15%-40%) and specificity (97%, 95% CI: 94%-98%) were lower. Nine months after training, RDT knowledge was high among staff involved with the public health management of measles (average quiz score of 80%) and was highest among those who received formal training (88%), followed by those trained during supervisory visits (83%). During the RDT implementation period, the number of days from case confirmation until initiation of public response decreased by about 5 days.

    CONCLUSION: The measles IgM RDT shows >95% inter-reader agreement, high retention of RDT knowledge, and a more rapid public health response. However, despite ≥95% RDT specificity using capillary blood or oral fluid, RDT sensitivity was <45%. Higher-powered studies using highly specific IgM assays and systematic RT-qPCR for case confirmation are needed to establish the role of RDT in measles elimination settings.