Displaying all 7 publications

Abstract:
Sort:
  1. Ahmad M, Roy RA
    Endod Dent Traumatol, 1994 Apr;10(2):71-6.
    PMID: 8062810
    The incidence of breakage of Piezon-Master ultrasonic K files were evaluated. Three groups of unused files were subjected to three treatments, namely; free vibration in air without irrigation, free vibration in root canal while minimizing contact with the wall of canal in the presence of irrigation and light filing in root canal with free flow of irrigation. Cavitation produced by files in contact and free of contact with a glass surface was examined in order to observe the relationship between cavitation defects and breakage. In addition, the fractured and unfractured files were examined under a scanning electron microscope for the presence of cavitation pits. The results indicated that more files broke in air. In water, a higher incidence of breakage occurred when files were allowed to freely vibrate while no breakage occurred when the files were used in filing. All files generated cavitation which resulted in pitting of their surfaces. However, it was considered unlikely that the pits contributed to fracture. Fatigue cracks which could be the result of the manufacturing process were observed at some of the corners of the cross sections of the fractured files and could be the main contributory factor to fracture.
  2. Lui JL
    Endod Dent Traumatol, 1992 Feb;8(1):36-8.
    PMID: 1396360
    A method of treating an apical root fracture with accompanying resorption at the junction of the fracture fragments using glass-cermet cement is described. Endodontically, the material had previously been used for repair of lateral resorptive root defects and retrograde root fillings. Complete bone regeneration was observed three years post-operatively following treatment of the root fracture in the conventional manner. The various advantages of glass-cermet cement as a root filling material used in the technique described are discussed.
  3. Ahmad M, Roy RA, Kamarudin AG
    Endod Dent Traumatol, 1992 Oct;8(5):189-94.
    PMID: 1302678
    The steady acoustic streaming generated around straight and precurved oscillating ultrasonic files driven by the Piezon-Master 400 unit was examined in the free field and in small channels using a stereomicroscope. In addition, the effect of file-wall contact on streaming production was also investigated. The results indicated that the ultrasonic files can generate acoustic streaming both in the free field and in the small channel. Higher velocity streaming was observed when smaller size files were employed and when the file was precurved. Light file-wall contact did not totally inhibit streaming while severe file-wall contact inhibited movement of the file and, as a result, no streaming was observed. The positions and length scales of the streaming vortices appeared to be influenced by the presence of boundaries. In the free field, two rows of vortices were situated along the sides of the file while in the small channel, the vortices were positioned above the surface of the file. These results indicated that it is possible for acoustic streaming to occur in a confined space as in a root canal provided that severe file-wall contact is avoided. It is therefore recommended that light filing or allowing the file to freely vibrate during some stage of treatment should be carried out in order to generate streaming in the root canal.
  4. Ahmad M
    Endod Dent Traumatol, 1991 Apr;7(2):55-8.
    PMID: 1782894
    The efficacy of two ultrasonic units in shaping curved canals in teeth were compared. Twenty teeth were instrumented using the Cavi-Endo unit at a power setting 1 using the technique recommended by the manufacturer. Another group of 20 teeth received similar treatment but were instrumented with the Enac unit. The time taken to instrument each canal was recorded. The pre- and post-instrumented radiographs of the teeth of x 10 magnification were taken using a microfocal technique. The radiographs were subjected to a subtraction technique to result in composite images of the pre- and post-instrumented shapes. The canal shape and the incidence of elbows were evaluated using various measurements taken from the radiographs. The manner the dentine was removed was similar in both groups. All canals exhibited unequal removal along the canal with more dentine being removed at the coronal end. The Enac group exhibited a higher incidence of elbows which occurred further apically than those in the Cavi-Endo group. There was no significant difference between groups in the following: time of instrumentation, amount of apical and coronal canal enlargement, apical deviation and change in width at the elbow. These findings were no different from those of another study using simulated canals.
  5. Ahmad M
    Endod Dent Traumatol, 1990 Feb;6(1):12-5.
    PMID: 2202589
    The resonant characteristics of ultrasonic files driven by the Cavi-Endo unit were examined. The investigation was carried out by evaluating the pattern of oscillation and the power emitted by the ultrasonic files. It was observed that the files oscillated in a sinusoidal fashion, exhibiting a standing wave pattern along the file. The power emitted was found to be directly proportional to the flexibility of the file. Such behaviour is typical of a system that exhibits resonance.
  6. Ahmad M
    Endod Dent Traumatol, 1990 Oct;6(5):230-1.
    PMID: 2133315
    The temperature rise of the irrigant in the root canal during free vibration of the ultrasonic file was studied in vitro in 10 human teeth. The mean temperature rise was found to be 0.6 degrees C. The minimal temperature increase may not significantly contribute to the effectiveness of ultrasonic root canal instrumentation.