  1. Piccini JP, Stromberg K, Jackson KP, Laager V, Duray GZ, El-Chami M, et al.
    Heart Rhythm, 2017 05;14(5):685-691.
    PMID: 28111349 DOI: 10.1016/j.hrthm.2017.01.026
    BACKGROUND: Device repositioning during Micra leadless pacemaker implantation may be required to achieve optimal pacing thresholds.

    OBJECTIVE: The purpose of this study was to describe the natural history of acute elevated Micra vs traditional transvenous lead thresholds.

    METHODS: Micra study VVI patients with threshold data (at 0.24 ms) at implant (n = 711) were compared with Capture study patients with de novo transvenous leads at 0.4 ms (n = 538). In both cohorts, high thresholds were defined as >1.0 V and very high as >1.5 V. Change in pacing threshold (0-6 months) with high (1.0 to ≤1.5 V) or very high (>1.5 V) thresholds were compared using the Wilcoxon signed-rank test.

    RESULTS: Of the 711 Micra patients, 83 (11.7%) had an implant threshold of >1.0 V at 0.24 ms. Of the 538 Capture patients, 50 (9.3%) had an implant threshold of >1.0 V at 0.40 ms. There were no significant differences in patient characteristics between those with and without an implant threshold of >1.0 V, with the exception of left ventricular ejection fraction in the Capture cohort (high vs low thresholds, 53% vs 58%; P = .011). Patients with an implant threshold of >1.0 V decreased significantly (P < .001) in both cohorts. Micra patients with high and very high thresholds decreased significantly (P < .01) by 1 month, with 87% and 85% having 6-month thresholds lower than the implant value. However, when the capture threshold at implant was >2 V, only 18.2% had a threshold of ≤1 V at 6 months and 45.5% had a capture threshold of >2 V.

    CONCLUSIONS: Pacing thresholds in most Micra patients with elevated thresholds decrease after implant. Micra device repositioning may not be necessary if the pacing threshold is ≤2 V.

    Matched MeSH terms: Bradycardia/therapy*
  2. Yew KL, Anderson S, Farah R, Lim SH
    Asian Cardiovasc Thorac Ann, 2014 Oct;22(8):979-80.
    PMID: 24887840 DOI: 10.1177/0218492313491583
    Central vein stenosis is not uncommon in hemodialysis-dependent patients as a result of mechanical damage to the vessel walls from prior cannulation. It can cause ipsilateral upper limb swelling and pain, resulting in suboptimal hemodialysis. It is unfortunate for bilateral central vein stenosis to develop concomitantly, and rare in the setting of an in-situ pacemaker. This case illustrates the successful ligation of a nondependent left arteriovenous fistula and stenting of the right subclavian vein with functioning ipsilateral arteriovenous fistula, to overcome the problem of symptomatic bilateral upper limb swelling.
    Matched MeSH terms: Bradycardia/therapy*