METHODS: We evaluated the performance characteristics of the LMA Protector™ in 30 unparalysed, moderately obese patients. First attempt insertion rate, time for insertion, oropharyngeal leak pressure (OLP), and incidence of complications were recorded.
RESULTS: We found high first and second attempt insertion rates of 28(93%) and 1(33%) respectively, with one failed attempt where no capnography trace could be detected, presumably from a downfolded device tip. The LMA Protector™ was inserted rapidly in 21.0(4.0) seconds and demonstrated high OLP of 31.8(5.4) cmH2O. Fibreoptic assessment showed a clear view of vocal cords in 93%. The incidence of blood staining on removal of device was 48%, postoperative sore throat 27%, dysphagia 10% and dysphonia 20% (all self-limiting, resolving a few hours postoperatively).
CONCLUSIONS: We conclude that the LMA Protector™ was associated with easy, expedient first attempt insertion success, demonstrating high oropharyngeal pressures and good anatomical position in the moderately obese population, with relatively low postoperative airway morbidity.
TRIAL REGISTRATION: Australian New Zealand Clinical Trials Registry, ACTRN12617001152314 . Registered 7 August 2017.