METHODS: This was an observational manikin-based study. A total of 96 participants as well as two types of mechanical compression devices: Lucas-2 and AutoPulse, performed one minute of continuous chest compression on BT-CPEA programmed manikin while the ambulance travelled at different speeds, i.e., idle state, 30km/hr and 60km/hr. Seven outcome variables of chest compression were measured. Performance data of different groups of compressor were compared and analysed using repeated measures analysis of variance (ANOVA).
RESULTS: In manual chest compression, significant variation were noted among different speeds in term of average compression rate (p<0.001), average compression depth (p=0.007), fraction of adequate/insufficient compression depth and fraction of normal hands positioning with p=0.018, 0.022 and 0.034 respectively. Overall, AutoPulse and Lucas-2 were not affected by ambulance speed. Lucas- 2 showed more consistent average compression rate, higher fraction of adequate compression depth and reduced fraction of insufficient compression depth as compared to manual compression with p<0.001, 0.001 and 0.043 respectively.
CONCLUSION: In this study we found that ambulance speed significantly affected certain aspects of manual chest compression most notably compression depth, rate and hand positioning. AutoPulse and Lucas-2 can improve these aspects by providing more consistent compression rate, depth and fraction of adequate compression depth during transport.