Affiliations 

  • 1 Department of Pharmacology, Faculty of Medicine, University of Malaya, 50603 Kuala Lumpur
Biomed Chromatogr, 2007 Aug;21(8):775-9.
PMID: 17497758

Abstract

A fast and simple capillary zone electrophoresis method was developed and validated for the determination of lidocaine in skin using tape samples. Separation was performed in a 350 mm (265 mm to window) x 50 microm i.d. fused silica capillary using a background electrolyte of phosphoric acid-Tris pH 2.5. The extraction of lidocaine from tape samples was achieved using methanol, which was diluted to 50% with water before injection. Procaine was the internal standard. The migration times for procaine and lidocaine were 2.9 and 3.2 min, respectively. The limit of quantification for lidocaine was 50 microg, with signal to noise ratio greater than 10. The calibration curve was linear from 50 to 1000 microg with r(2) greater than 0.99. The CV for both within- and between-assay imprecision and the percentage of inaccuracy for the quality control samples including lower and upper limits of quantitation were 97%. The accuracy and selectivity of this method allowed the measurement of lidocaine in tape samples obtained from a skin tape stripping study of local anesthetics in healthy subjects.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.