Affiliations 

  • 1 Department of Medical Microbiology and Immunology, Faculty of Medicine, Universiti Kebangsaan Malaysia
Malays J Pathol, 1992 Dec;14(2):105-10.
PMID: 1338997

Abstract

The group A rotavirus staphylococcal co-agglutination test was evaluated and its sensitivity and specificity compared with an in-house enzyme-linked immunosorbent assay (ELISA) and a commercial latex agglutination test (Rotalex). In addition, the storage stability of the staphylococcal reagents was ascertained. Examination of 136 clarified suspensions of diarrhoeal faeces by the staphylococcal co-agglutination test revealed a high proportion of false positives (26%) and uninterpretable results (34%) due to non-specific agglutination. Non-specific agglutination could be removed effectively by prior absorption of the clarified faecal specimens with unsensitized staphylococci. The staphylococcal co-agglutination test was less sensitive and specific than the in-house enzyme-linked immunosorbent assay but was comparable to the Rotalex slide latex agglutination test. The staphylococcal reagents have a shelf life of at least 29 weeks.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.