Peripheral vestibular disorder (PVD) is serious and common. Clinically, giving an accurate diagnosis of PVD can be challenging. Vestibular evoked myogenic potential (VEMP) is an objective test to evaluate the integrity of vestibular organs, particularly saccule and/or inferior vestibular nerve. This study was performed to determine the sensitivity and specificity of VEMP using different stimuli. Fourty normal and 65 PVD subjects who fulfilled the inclusion criteria were recruited. While sitting comfortably, VEMP waveforms were recorded with active electrode on sternocleidomastoid muscle and negative electrode on upper forehead. Tone bursts (500, 750 and 1000 Hz) were delivered via headphones at 90 dBnHL and 5/s presentation rate. VEMP parameters for each stimulus (amplitude and latency of P1 and N1 peak) were analyzed accordingly. Receiver operating characteristic (ROC) was performed to determine the sensitivity and specificity of VEMP at different test frequencies. N1 amplitude of 750 Hz stimulus produced the most ideal sensitivity (65% on right and 63% on left) and specificity (83% on right and 78% on left). The importance of using a few tone bursts in VEMP test in order to minimize the false negative in cases might be encountered in clinics as the certain tone burst had inadequate sensitivity in detecting PVD cases. The 750 Hz stimulus produced the most ideal VEMP with adequate values of sensitivity and specificity, at least in this study.