Affiliations 

  • 1 Centre for Applied Biomechanics, Department of Biomedical Engineering, Faculty of Engineering, University of Malaya, Kuala Lumpur, Malaysia
Proc Inst Mech Eng H, 2020 Aug;234(8):884-894.
PMID: 32459140 DOI: 10.1177/0954411920923541

Abstract

Many studies have shown that medical compression products produce different levels of interface pressure during the usage of the products. However, limited studies have explored the pattern of interface pressure exerted by orthotic garments. This case study aimed to investigate the pattern of interface pressure exerted by two types of orthotic garments on a child with cerebral palsy. A 13-year-old child diagnosed with ataxic spastic diplegia cerebral palsy has difficulty to perform sit-to-stand motion even with a walking frame due to his truncal ataxia. A TheraTogsTM orthosis and a Dynamic Lycra® Fabric Orthosis (DLFO) were prepared for the child. The child's sit-to-stand ability without and with the usage of orthoses was recorded using five sit-to-stand tests. The garments' interface pressure was measured using F-scan (9811E) and F-scan 6.5.1 version software. The pressure was recorded when the child was in sitting position and performing sit-to-stand-to-sit motion. Overall, the child completed the five sit-to-stand test duration within 2.53 ± 0.04 s and 2.51 ± 0.09 s with the usage of TheraTogsTM orthosis and DLFO, respectively. Higher pressure was exerted by Dynamic Lycra Fabric Orthosis (axillary = 122 mmHg) in contrast to TheraTogsTM orthosis (77 mmHg) when the child was in a sitting position. Lower pressure was exerted by DLFO (7 mmHg), over xiphoid level and for TheraTogsTM orthosis is 1.2 mmHg over axillary level when the child was performing sit-to-stand motion. The largest range of pressure was exerted by TheraTogsTM orthosis with a minimum pressure of 5 mmHg and a maximum pressure of 155 mmHg during sit-to-stand motion. Overall, the DLFO exerted higher mean interface pressure on the child in comparison to TheraTogsTM orthosis when the child's body was in a sitting position wearing both upper garment and pants. Both TheraTogsTM orthosis and DLFO presented a different range of interface pressure over different body segments and activities.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.