Breast cancer is the most frequently occurring cancer in women globally and early detection increases the survival rate of patients. Therefore, this study was done to determine factors which influence the awareness of breast cancer and practice of screening procedures. A cross-sectional study was performed on 125 women aged 19-60 years in urban and rural areas in Malaysia using a validated questionnaire covering knowledge of breast cancer and screening practices. A total of 99.2% respondents knew that breast cancer is the leading cancer with a mean knowledge of 67.3 ± 15.3% for urban and 50.2 ± 14.7% for rural women Mann Whitney U showed rural women had significantly less awareness compared to urban women (p< 0.05). Spearman correlation test showed a significant positive relationship between education and awareness (p< 0.05). Regarding awareness of the screening methods, 92.8%, 50.4% and 47.2% of respondents correctly answered questions on capability of BSE, CBE and mammography, respectively. In conclusion, the study showed awareness of breast cancer and practice of screening procedures increases with higher education and urban living. Therefore, there is an urgent need for an intensive breast cancer awareness campaign and availablity of screening centres prioritized in rural areas.
BACKGROUND: Optimisation of average glandular dose (AGD) for two-dimensional (2D) mammography is important, as imaging using ionizing radiation has the probability to induce cancer resulting from stochastic effects. This study aims to observe the effects of kVp, anode/filter material, and exposure mode on the dose and image quality of 2D mammography.
METHODS: This experimental study was conducted using full-field digital mammography. The entrance surface air kerma was determined using thermoluminescent dosimeter (TLD) 100H and ionization chamber (IC) on three types of Computerized Imaging Reference System (CIRS) phantom with 50/50, 30/70, and 20/80 breast glandularity, respectively, in the auto-time mode and auto-filter mode. The Euref protocol was used to calculate the AGD while the image quality was evaluated using contrast-to-noise ratio (CNR), figure of merit (FOM), and image quality figure (IQF).
RESULTS: It is shown that AGD values in the auto-time mode did not decrease significantly with the increasing tube voltage of the silver filter (r = -0.187, P > 0.05) and rhodium filter (r = -0.131, P > 0.05) for all the phantoms. The general linear model showed that AGD for all phantoms had a significant effect between different exposure factors [F (6,12.3) = 4.48 and mode of exposure F (1,86) = 4.17, P < 0.05, respectively] but there is no significant difference between the different anode/filter combination [F (1,4) = 0.571].
CONCLUSION: In summary, the 28, 29, and 31 kVp are the optimum kVp for 50%, 30%, and 20% breast glandularity, respectively. Besides the auto-filter mode is suitable for 50%, 30%, and 20% breast glandularity because it is automatic, faster, and may avoid error done by the operator.
KEYWORDS: CDMAM; digital mammography; radiation dose
Full field digital mammography (FFDM) has been progressively introduced in medical centers in recent years. However, it is questionable which exposure parameters are suitable in order to reduce the glandular breast doses as they are related to induced carcinogenesis. The goal of this study was to compare the average glandular doses (AGD) and image quality of three FFDM systems namely Siemens Mammomat NovationDR, Hologic Lorad Selenia and General Electric Senographe Essential using a Figure of Merit. A Computerized Imaging Reference Systems (CIRS) tissue equivalent breast phantom which consists of phototimer compensation plate with different thickness and glandularity was exposed in fully automatic exposure control mode in the cranio-caudal projection similar to clinical settings. Thermoluminescent dosimeter 100H (TLD- 100H) was used to measure the entrance surface air kerma (ESAK), the AGD was calculated using European protocol whilst the image quality was assessed quantitatively by measuring the contrast to noise ratio (CNR) value. The obtained values were used to calculate the Figure of Merit (FOM) to analyze the effectiveness of the system. Repeated Measures ANOVA analysis showed that there is a significant difference (p<0.05) in the mean value of AGD and CNR between the three FFDM systems. Hologic Lorad Selenia system contrbuted the highest AGD value while General Electric Senographe Essential had the highest CNR and FOM value. In conclusion, this study may provide an objective criterion during the selection of a mammography unit by using the figure of merit for screening or diagnostic purpose.