Affiliations 

  • 1 Department Nuclear Medicine, Radiotherapy and Oncology, School of Medical Sciences, Universiti Sains Malaysia Health Campus, 16150 Kubang Kerian, Kelantan, Malaysia
Radiat Prot Dosimetry, 2013;154(3):396-9.
PMID: 23012482 DOI: 10.1093/rpd/ncs239

Abstract

A femur phantom made of wax and a real human bone was used to study the dose during radiographical procedures. The depth dose inside the phantom was determined using DOSXYZnrc, a Monte Carlo simulation software. The results were verified with measurements using TLD-100H. It was found that for 2.5 mm aluminium filtered 84-kVp X-rays, the radiation dose in the bone reached 57 % higher than the surface dose, i.e. 3.23 mGy as opposed to 2.06 mGy at the surface. The use of real bone introduces variations in the bone density in the DOSXYZnrc model, resulting in a lower attenuation effect than expected from solid bone tissues.

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.