Affiliations 

  • 1 Department of Radiology, Universiti Malaysia Sarawak, Kuching, Malaysia
Biomed Imaging Interv J, 2012 Jan;8(1):e8.
PMID: 22970064 MyJurnal DOI: 10.2349/biij.8.1.e8

Abstract

Fibrin sheath formation around long-term haemodialysis catheter is a common cause of failed dialysis access. Treatment options include pharmacological and mechanical methods. This paper reports a case of failed dialysis access due to fibrin sheath encasement. Pharmacologic thrombolysis, mechanical disruption using guide wire and catheter exchange had failed to address the issue. Eventually, fibrin sheath stripping using the loop snare technique was able to successfully restore the catheter function.