A simple high-performance liquid chromatographic method was developed for the determination of omeprazole in human plasma. Omeprazole and the internal standard, chloramphenicol, were extracted from alkalinized plasma samples using dichloromethane. The mobile phase was 0.05 M Na2HPO4-ACN (65:35, v/v) adjusted to pH 6.5. Analysis was run at a flow rate of 1.0 ml/min at a detection wavelength of 302 nm. The method was specific and sensitive with a detection limit of 2.5 ng/ml at a signal-to-noise ratio of 4:1. The limit of quantification was set at 5 ng/ml. The calibration curve was linear over a concentration range of 5-1280 ng/ml. Mean recovery value of the extraction procedure was about 96%, while the within and between day coefficient of variation and percent error values of the assay method were all less than 14%.