Ultrasound measurement of ocular dimension is the chosen method for assessing axial length when determining dioptric power for intraocular lens. From the current results of 30 cases studied, the mean axial length ranges from 22 to 23 mm. Despite the limitation of the accuracy of the ultrasonic measurements with the 7.5 mHz transducer, the power of intraocular lens can. be determined satisfactorily in accordance with the knowledge of keratometric reading. Hence, high refractive errors could be avoided post-operatively.