Displaying publications 1 - 20 of 79 in total

Abstract:
Sort:
  1. Jamil A, Mohd MI, Zain NM
    Radiat Prot Dosimetry, 2018 Dec 01;182(4):413-418.
    PMID: 29767799 DOI: 10.1093/rpd/ncy082
    After years of establishment of computed radiography (CR) and digital radiography (DR), manufacturers have introduced exposure indicator/index (EI) as a feedback mechanism for patient dose. However, EI consistency is uncertain for CR. Most manufacturers recommended EI values in a range of numbers for all examination, instead of giving the exact range for a specific body part, raising a concern of inappropriate exposure given to the patient in clinical practice. The aims of this study were to investigate the EI consistency in DR systems produced in constant exposure parameters and clinical condition, and to determine the interaction between the anatomical part and EI. A phantom study of skull, chest, abdomen and hand was carried out and four systems were used for comparison-Fuji CR, Carestream CR, Siemens DR and Carestream DR. For each projection, the phantom positioning and exposure parameters were set according to the standard clinical practice. All exposure parameters and clinical conditions were kept constant. Twenty (20) exposures were taken for each projection and the EI was recorded. Findings showed that EI is not consistent in DR systems despite constant exposure parameters and clinical condition except in Siemens DR, through skull examination. Statistical analysis showed a significant interaction between anatomical parts and EI values (P < 0.05). EI alone was proven to be less reliable to provide technologist a correct feedback on exposure level. The interaction between anatomical parts and EI values intensifies the need for an anatomical-specific EI values set by all manufacturers for accurate feedback on the exposure parameters used and the detector entrance dose.
    Matched MeSH terms: Radiometry/instrumentation*; Radiometry/standards*
  2. Salehi Z, Yusoff AL
    Radiat Prot Dosimetry, 2013;154(3):396-9.
    PMID: 23012482 DOI: 10.1093/rpd/ncs239
    A femur phantom made of wax and a real human bone was used to study the dose during radiographical procedures. The depth dose inside the phantom was determined using DOSXYZnrc, a Monte Carlo simulation software. The results were verified with measurements using TLD-100H. It was found that for 2.5 mm aluminium filtered 84-kVp X-rays, the radiation dose in the bone reached 57 % higher than the surface dose, i.e. 3.23 mGy as opposed to 2.06 mGy at the surface. The use of real bone introduces variations in the bone density in the DOSXYZnrc model, resulting in a lower attenuation effect than expected from solid bone tissues.
    Matched MeSH terms: Radiometry/methods*
  3. Samson DO, Aziz MZA, Shukri A, Mat Jafri MZ, Hashim R, Zuber SH, et al.
    Health Phys, 2023 Aug 01;125(2):77-91.
    PMID: 36826380 DOI: 10.1097/HP.0000000000001688
    The current study was undertaken to investigate the radiological and dosimetric parameters of natural product-based composite (SPI/NaOH/IA-PAE/ Rhizophora spp .) phantoms. The radiological properties of the phantoms were measured at different gamma energies from Compton scatter of photons through angles of 0, 30, 45, 60, 75, and 90 degrees. Ionization chamber (IC) and Gafchromic EBT3 film dosimeters were employed to evaluate the dosimetric characteristics for photons (6-10 MV) and electrons (6-15 MeV). Radiological property results of the composite phantoms were consistent with good quality compared to those of solid water phantoms and theoretical values of water. Photon beam quality index of the SPI15 phantom with p-values of 0.071 and 0.073 exhibited insignificant changes. In addition, good agreement was found between PDD curves measured with IC and Gafchromic EBT3 film for both photons and electrons. The computed therapeutic and half-value depth ranges matched within the limits and are similar to those of water and solid water phantoms. Therefore, the radiological and dosimetric parameters of the studied composite phantom permit its use in the selection of convenient tissue- and water-equivalent phantom material for medical applications.
    Matched MeSH terms: Radiometry/methods
  4. Jong WL, Ung NM, Vannyat A, Rosenfeld AB, Wong JHD
    Phys Med, 2017 Oct;42:39-46.
    PMID: 29173919 DOI: 10.1016/j.ejmp.2017.08.011
    Challenges in treating lung tumours are related to the respiratory-induced tumour motion and the accuracy of dose calculation in charged particle disequilibrium condition. The dosimetric characteristics near the interface of lung and Perspex media in a moving phantom during respiratory-gated and non-gated radiotherapy were investigated using Gafchromic EBT2 and the MOSkin detector. The MOSkin detectors showed good agreement with the EBT2 films during static and gated radiotherapy. In static radiotherapy, the penumbral widths were found to be 3.66mm and 7.22mm in Perspex and lung media, respectively. In non-gated (moving) radiotherapy with 40mm respiratory amplitude, dose smearing effect was observed and the penumbral widths were increased to 28.81mm and 26.40mm, respectively. This has been reduced to 6.85mm and 9.81mm, respectively, in gated radiotherapy with 25% gating window. There were still some dose discrepancies as compared to static radiotherapy due to the residual motion. This should be taken into account in the margin generation for the target tumour.
    Matched MeSH terms: Radiometry/instrumentation
  5. Wahabi JM, Ung NM, Mahdiraji GA, Wong JHD
    Radiat Prot Dosimetry, 2024 Mar 02;200(3):264-273.
    PMID: 38123475 DOI: 10.1093/rpd/ncad303
    The radioluminescent (RL) dosemeter is excellent for real-time radiation measurement and can be used in various applications. A plastic scintillator is often the choice sensor because of its size and tissue equivalency. This study aims to characterise a novel plastic scintillator irradiated with high-energy photon beams. An RL dosimetry system was developed using the plastic scintillator. The RL dosimetry system was irradiated using a linear accelerator to characterise the dose linearity, dose rate, energy dependency and depth dose. The developed system showed a linear response toward the dose and dose rate. An energy dependency factor of 1.06 was observed. Depth dose measurement showed a mean deviation of 1.21% from the treatment planning system. The response and characteristics of the plastic scintillator show that it may be used as an alternative in an RL dosimetry system.
    Matched MeSH terms: Radiometry*
  6. Jong WL, Ung NM, Vannyat A, Jamalludin Z, Rosenfeld A, Wong JH
    Phys Med, 2017 Jan;33:127-135.
    PMID: 28089106 DOI: 10.1016/j.ejmp.2016.12.020
    Dosimetry in small radiation field is challenging and complicated because of dose volume averaging and beam perturbations in a detector. We evaluated the suitability of the "Edge-on" MOSkin (MOSFET) detector in small radiation field measurement. We also tested the feasibility for dosimetric verification in stereotactic radiosurgery (SRS) and stereotactic radiotherapy (SRT). "Edge-on" MOSkin detector was calibrated and the reproducibility and linearity were determined. Lateral dose profiles and output factors were measured using the "Edge-on" MOSkin detector, ionization chamber, SRS diode and EBT2 film. Dosimetric verification was carried out on two SRS and five SRT plans. In dose profile measurements, the "Edge-on" MOSkin measurements concurred with EBT2 film measurements. It showed full width at half maximum of the dose profile with average difference of 0.11mm and penumbral width with difference of ±0.2mm for all SRS cones as compared to EBT2 film measurement. For output factor measurements, a 1.1% difference was observed between the "Edge-on" MOSkin detector and EBT2 film for 4mm SRS cone. The "Edge-on" MOSkin detector provided reproducible measurements for dose verification in real-time. The measured doses concurred with the calculated dose for SRS (within 1%) and SRT (within 3%). A set of output correction factors for the "Edge-on" MOSkin detector for small radiation fields were derived from EBT2 film measurement and presented. This study showed that the "Edge-on" MOSkin detector is a suitable tool for dose verification in small radiation field.
    Matched MeSH terms: Radiometry/instrumentation*
  7. Jamalludin Z, Jong WL, Malik RA, Rosenfeld AB, Ung NM
    Phys Med, 2020 Jan;69:52-60.
    PMID: 31830631 DOI: 10.1016/j.ejmp.2019.11.025
    PURPOSE: Dose to the rectum during brachytherapy treatment may differ from an approved treatment plan which can be quantified with in vivo dosimetry (IVD). This study compares the planned with in vivo doses measured with MOSkin and PTW 9112 rectal probe in patients undergoing CT based HDR cervical brachytherapy with Co-60 source.

    METHODS: Dose measurement of a standard pear-shaped plan carried out in phantom to verify the MOSkin dose measurement accuracy. With MOSkin attached to the third diode, RP3 of the PTW 9112, both detectors were inserted into patients' rectum. The RP3 and MOSkin measured doses in 18 sessions as well as the maximum measured doses from PTW 9112, RPmax in 48 sessions were compared to the planned doses.

    RESULTS: Percentage dose differences ΔD (%) in phantom study for two MOSkin found to be 2.22 ± 0.07% and 2.5 ± 0.07%. IVD of 18 sessions resulted in ΔD(%) of -16.3% to 14.9% with MOSkin and ΔD(%) of -35.7% to -2.1% with RP3. In 48 sessions, RPmax recorded ΔD(%) of -37.1% to 11.0%. MOSkin_measured doses were higher in 44.4% (8/18) sessions, while RP3_measured were lower than planned doses in all sessions. RPmax_measured were lower in 87.5% of applications (42/47).

    CONCLUSIONS: The delivered doses proven to deviate from planned doses due to unavoidable shift between imaging and treatment as measured with MOSkin and PTW 9112 detectors. The integration of MOSkin on commercial PTW 9112 surface found to be feasible for rectal dose IVD during cervical HDR ICBT.

    Matched MeSH terms: Radiometry
  8. Jamalludin Z, Jong WL, Abdul Malik R, Rosenfeld A, Ung NM
    Phys Med, 2019 Feb;58:1-7.
    PMID: 30824140 DOI: 10.1016/j.ejmp.2019.01.010
    In vivo dosimetry in high dose-rate (HDR) intracavitary brachytherapy (ICBT) is important for assessing the true dose received by surrounding organs at risk during treatment. It also serves as part of the treatment delivery quality assurance and verification program with the use of a suitable dosimeter. Such a dosimeter should be characterized under brachytherapy conditions before clinical application to ensure the accuracy of in vivo measurement. In this study, a MOSFET-based detector, MOSkin, was calibrated and characterized under HDR Cobalt-60 (Co-60) brachytherapy source. MOSkin possessed the major advantages of having small physical and dosimetric sizes of 4.8 × 10-6 mm3 with the ability to provide real-time measurements. Using solid water and polymethyl methacrylate (PMMA) phantom, the detectors' reproducibility, linearity, angular and distance dependency was tested for its suitability as an in vivo detector. Correction factors to account for differences in depth measurements were determined. The MOSkin detector showed a reliable response when tested under Co-60 brachytherapy range of doses with an excellent linearity of R2 = 0.9997 and acceptable reproducibility. A phantom verification study was also conducted to verify the differences between MOSkin responses and treatment planning (TPS) calculated doses. By taking into account several correction factors, deviations ranging between 0.01 and 0.4 Gy were found between MOSkin measured and TPS doses at measurement distance of 20-55 mm. The use of MOSkin as the dosimeter of choice for in vivo dosimetry under Co-60 brachytherapy condition is feasible.
    Matched MeSH terms: Radiometry/instrumentation*
  9. Kodaira S, Konishi T, Kobayashi A, Maeda T, Ahmad TA, Yang G, et al.
    J Radiat Res, 2015 Mar;56(2):360-5.
    PMID: 25324538 DOI: 10.1093/jrr/rru091
    The geometric locations of ion traversals in mammalian cells constitute important information in the study of heavy ion-induced biological effect. Single ion traversal through a cellular nucleus produces complex and massive DNA damage at a nanometer level, leading to cell inactivation, mutations and transformation. We present a novel approach that uses a fluorescent nuclear track detector (FNTD) for the simultaneous detection of the geometrical images of ion traversals and DNA damage in single cells using confocal microscopy. HT1080 or HT1080-53BP1-GFP cells were cultured on the surface of a FNTD and exposed to 5.1-MeV/n neon ions. The positions of the ion traversals were obtained as fluorescent images of a FNTD. Localized DNA damage in cells was identified as fluorescent spots of γ-H2AX or 53BP1-GFP. These track images and images of damaged DNA were obtained in a short time using a confocal laser scanning microscope. The geometrical distribution of DNA damage indicated by fluorescent γ-H2AX spots in fixed cells or fluorescent 53BP1-GFP spots in living cells was found to correlate well with the distribution of the ion traversals. This method will be useful for evaluating the number of ion hits on individual cells, not only for micro-beam but also for random-beam experiments.
    Matched MeSH terms: Radiometry/instrumentation*
  10. Mizuno H, Fukuda S, Fukumura A, Nakamura YK, Jianping C, Cho CK, et al.
    J Radiat Res, 2017 May 01;58(3):372-377.
    PMID: 27864507 DOI: 10.1093/jrr/rrw108
    A dose audit of 16 facilities in 11 countries has been performed within the framework of the Forum for Nuclear Cooperation in Asia (FNCA) quality assurance program. The quality of radiation dosimetry varies because of the large variation in radiation therapy among the participating countries. One of the most important aspects of international multicentre clinical trials is uniformity of absolute dose between centres. The National Institute of Radiological Sciences (NIRS) in Japan has conducted a dose audit of participating countries since 2006 by using radiophotoluminescent glass dosimeters (RGDs). RGDs have been successfully applied to a domestic postal dose audit in Japan. The authors used the same audit system to perform a dose audit of the FNCA countries. The average and standard deviation of the relative deviation between the measured and intended dose among 46 beams was 0.4% and 1.5% (k = 1), respectively. This is an excellent level of uniformity for the multicountry data. However, of the 46 beams measured, a single beam exceeded the permitted tolerance level of ±5%. We investigated the cause for this and solved the problem. This event highlights the importance of external audits in radiation therapy.
    Matched MeSH terms: Radiometry*
  11. Al-jarrah AM, Abdul Rahman A, Shahrim I, Razak NN, Ababneh B, Tousi ET
    Phys Med, 2016 Jan;32(1):36-41.
    PMID: 26494156 DOI: 10.1016/j.ejmp.2015.09.003
    Genipin gel dosimeters are hydrogels infused with a radiation-sensitive material which yield dosimetric information in three dimensions (3D). The effect of inorganic salts and glucose on the visible absorption dose-response, melting points and mass density of genipin gel dosimeters has been experimentally evaluated using 6-MV LINAC photons. As a result, the addition of glucose with optimum concentration of 10% (w/w) was found to improve the thermal stability of the genipin gel and increase its melting point (Tm) by 6 °C accompanied by a slight decrease of dose-response. Furthermore, glucose helps to adjust the gel mass density to obtain the desired tissue-equivalent properties. A drop of Tm was observed when salts were used as additives. As the salt concentration increased, gel Tm decreased. The mass density and melting point of the genipin gel could be adjusted using different amounts of glucose that improved the genipin gel suitability for 3D dose measurements without introducing additional toxicity to the final gel.
    Matched MeSH terms: Radiometry/methods*
  12. Cheah SK, Matthews T, Teh BS
    Asian Pac J Cancer Prev, 2016;17(9):4233-4235.
    PMID: 27797223
    BACKGROUND: Whole brain radiotherapy (WBRT) and stereotactic radiosurgery were frequently used to palliate patients with brain metastases. It remains controversial which modality or combination of therapy is superior especially in the setting of limited number of brain metastases. The availability of newer medical therapy that improves survival highlighted the importance of reducing long term radiation toxicity associated with WBRT. In this study, we aim to demonstrate the hippocampal sparing technique with whole brain and integrated simultaneous boost Materials and Methods: Planning data from 10 patients with 1-5 brain metastases treated with SRS were identified. Based on the contouring guideline from RTOG atlas, we identified and contoured the hippocampus with 5mm isocentric expansion to form the hippocampal avoidance structure. The plan was to deliver hippocampal sparing whole brain radiotherapy (HSWBRT) of 30 Gy in 10 fractions and simultaneous boost to metastatic lesions of 30 Gy in 10 fractions each.

    RESULTS: The PTV, hippocampus and hippocampal avoidance volumes ranges between 1.00 - 39.00 cc., 2.50 - 5.30 cc and 26.47 - 36.30 cc respectively. The mean hippocampus dose for the HSWBRT and HSWBRT and SIB plans was 8.06 Gy and 12.47 respectively. The max dose of optic nerve, optic chiasm and brainstem were kept below acceptable range of 37.5 Gy.

    CONCLUSIONS: The findings from this dosimetric study demonstrated the feasibility and safety of treating limited brain metastases with HSWBRT and SIB. It is possible to achieve the best of both worlds by combining HSWBRT and SIB to achieve maximal local intracranial control while maintaining as low a dose as possible to the hippocampus thereby preserving memory and quality of life.

    Matched MeSH terms: Radiometry
  13. Fum WKS, Wong JHD, Tan LK
    Phys Med, 2021 Apr;84:228-240.
    PMID: 33849785 DOI: 10.1016/j.ejmp.2021.03.004
    PURPOSE: This systematic review aims to understand the dose estimation approaches and their major challenges. Specifically, we focused on state-of-the-art Monte Carlo (MC) methods in fluoroscopy-guided interventional procedures.

    METHODS: All relevant studies were identified through keyword searches in electronic databases from inception until September 2020. The searched publications were reviewed, categorised and analysed based on their respective methodology.

    RESULTS: Hundred and one publications were identified which utilised existing MC-based applications/programs or customised MC simulations. Two outstanding challenges were identified that contribute to uncertainties in the virtual simulation reconstruction. The first challenge involves the use of anatomical models to represent individuals. Currently, phantom libraries best balance the needs of clinical practicality with those of specificity. However, mismatches of anatomical variations including body size and organ shape can create significant discrepancies in dose estimations. The second challenge is that the exact positioning of the patient relative to the beam is generally unknown. Most dose prediction models assume the patient is located centrally on the examination couch, which can lead to significant errors.

    CONCLUSION: The continuing rise of computing power suggests a near future where MC methods become practical for routine clinical dosimetry. Dynamic, deformable phantoms help to improve patient specificity, but at present are only limited to adjustment of gross body volume. Dynamic internal organ displacement or reshaping is likely the next logical frontier. Image-based alignment is probably the most promising solution to enable this, but it must be automated to be clinically practical.

    Matched MeSH terms: Radiometry
  14. Ilyas M, Abdul Aziz D, Tajuddin MR
    Int J Dermatol, 1988 Jun;27(5):315-8.
    PMID: 3391727
    Results from a 6-year study of solar ultraviolet A (UVA) radiation measurements at the equatorial location of Penang (5 degrees N) are presented. On clear days, the diurnal flux reaches a very high dosage of about 3.0 x 10(-2) KWHM-2 around midday. The average daily total flux is in the range of 1.6 x 10(-1) KWHM-2 and does not change much seasonally. The high 83% cloud cover only reduces the incoming flux to about half. The radiation flux represents a lower limit of the incident UVA radiation applicable to much of the equatorial/tropical region.
    Matched MeSH terms: Radiometry/methods*
  15. Sabarudin A, Md Yusof AK, Tay MF, Ng KH, Sun Z
    Radiat Prot Dosimetry, 2013;153(4):441-7.
    PMID: 22807493 DOI: 10.1093/rpd/ncs127
    This study was conducted to investigate the effectiveness of dose-saving protocols in dual-source computed tomography (CT) coronary angiography compared with invasive coronary angiography (ICA). On 50 patients who underwent coronary CT angiography was performed dual-source CT (DSCT) and compared with ICA procedures. Entrance skin dose (ESD), which was measured at the thyroid gland, and effective dose (E) were assessed for both imaging modalities. The mean ESD measured at the thyroid gland was the highest at 120 kVp, followed by the 100 kVp DSCT and the ICA protocols with 4.0±1.8, 2.7±1.0 and 1.1±1.2 mGy, respectively. The mean E was estimated to be 10.3±2.1, 6.2±2.3 and 5.3±3.4 mSv corresponding to the 120-kVp, 100-kVp DSCT and ICA protocols, respectively. The application of 100 kVp in DSCT coronary angiography is feasible only in patients with a low body mass index of <25 kg m(-2), which leads to a significant dose reduction with the radiation dose being equivalent to that of ICA.
    Matched MeSH terms: Radiometry/methods
  16. Tan SK, Yeong CH, Ng KH, Abdul Aziz YF, Sun Z
    PLoS One, 2016;11(8):e0161543.
    PMID: 27552224 DOI: 10.1371/journal.pone.0161543
    OBJECTIVES: This study aimed to measure the absorbed doses in selected organs for prospectively ECG-triggered coronary computed tomography angiography (CCTA) using five different generations CT scanners in a female adult anthropomorphic phantom and to estimate the effective dose (HE).

    MATERIALS AND METHODS: Prospectively ECG-triggered CCTA was performed using five commercially available CT scanners: 64-detector-row single source CT (SSCT), 2 × 32-detector-row-dual source CT (DSCT), 2 × 64-detector-row DSCT and 320-detector-row SSCT scanners. Absorbed doses were measured in 34 organs using pre-calibrated optically stimulated luminescence dosimeters (OSLDs) placed inside a standard female adult anthropomorphic phantom. HE was calculated from the measured organ doses and compared to the HE derived from the air kerma-length product (PKL) using the conversion coefficient of 0.014 mSv∙mGy-1∙cm-1 for the chest region.

    RESULTS: Both breasts and lungs received the highest radiation dose during CCTA examination. The highest HE was received from 2 × 32-detector-row DSCT scanner (6.06 ± 0.72 mSv), followed by 64-detector-row SSCT (5.60 ± 0.68 and 5.02 ± 0.73 mSv), 2 × 64-detector-row DSCT (1.88 ± 0.25 mSv) and 320-detector-row SSCT (1.34 ± 0.48 mSv) scanners. HE calculated from the measured organ doses were about 38 to 53% higher than the HE derived from the PKL-to-HE conversion factor.

    CONCLUSION: The radiation doses received from a prospectively ECG-triggered CCTA are relatively small and are depending on the scanner technology and imaging protocols. HE as low as 1.34 and 1.88 mSv can be achieved in prospectively ECG-triggered CCTA using 320-detector-row SSCT and 2 × 64-detector-row DSCT scanners.

    Matched MeSH terms: Radiometry
  17. Shakhreet BZ, Bauk S, Tajuddin AA, Shukri A
    Radiat Prot Dosimetry, 2009 Jul;135(1):47-53.
    PMID: 19482883 DOI: 10.1093/rpd/ncp096
    The mass attenuation coefficients (mu/rho) of Rhizophora spp. were determined for photons in the energy range of 15.77-25.27 keV. This was carried out by studying the attenuation of X-ray fluorescent photons from zirconium, molybdenum, palladium, silver, indium and tin targets. The results were compared with theoretical values for average breast tissues in young-age, middle-age and old-age groups calculated using photon cross section database (XCOM), the well-known code for calculating attenuation coefficients and interaction cross-sections. The measured mass attenuation coefficients were found to be very close to the calculated XCOM values in breasts of young-age group.
    Matched MeSH terms: Radiometry/methods*
  18. Banjade DP, Tajuddin AA, Shukri A
    Appl Radiat Isot, 2001 Sep;55(3):297-302.
    PMID: 11515650
    Previous scattering and depth-dose investigations involving use of the Malaysian hardwood Rhizophora spp have shown this medium to produce good agreement with measurements made in water. Present study extends the comparison, now including measurements of percentage depth-dose made for photons at 6MV and 5 and 12MeV electron beams. For the 6 MV photon and 5 MeV electron beams, discrepancies between percentage depth-dose for Rhizophora spp and water, at all depths, are found to be within 2.6 and 2.4% respectively. At 12 MeV electron energies, measured percentage depth-doses in Rhizophora spp beyond 3.5cm depth are found to be in significant discord with those for water. The absorbed dose in water measured in Rhizophora spp at d(max) for all three beams produces discrepancies of no more than 1.1% when compared with measurements made in water.
    Matched MeSH terms: Radiometry/methods*
  19. Banjade DP, Tajuddin AA, Shukri A
    Appl Radiat Isot, 2001 Aug;55(2):235-43.
    PMID: 11393765
    Protocols developed for high-energy dosimetry IAEA (Technical Reports Series No. 277, 1997), AAPM (Med. Phys. 10 (1983) 741: Med. Phys. 18 (1991) 73: Med. Phys. 21 (1994) 1251), IPEMB (Phys. Med. Biol. 41 (1996) 2557), and HPA (Phys. Med. Biol. 28 (1983) 1097) have continued to enhance precision in dose measurements and the optimization of radiotherapy procedures. While recent dosimetry protocols, including those due to the IAEA and IPEMB, have made a number of improvements compared with previous protocols, it is further desirable to develop absolute dosimetry methods of dose measurements. Measurements based on careful implementation of procedures contained within the various protocols have been carried out in an effort to determine the extent to which discrepancies exist among the protocols. Dose in water at dmax was measured using cylindrical and parallel-plate ionization chambers for 6 MV photon beams and 5 and 12 MeV electron beams. Results obtained from the use of the AAPM and HPA protocols for 6 MV photon beams were found to be 0.9% larger and 0.1% smaller, respectively, than those measured following the IAEA protocol. Calibration dose measurements for 5 and 12 MeV electron beams in water phantoms were found to agree to within 1%, this being well within recommendations from the ICRU and other sources regarding the accuracy of dose delivery.
    Matched MeSH terms: Radiometry/instrumentation; Radiometry/methods*; Radiometry/standards
  20. Banjade DP, Raj TA, Ng BS, Xavier S, Tajuddin AA, Shukri A
    Med Dosim, 2003;28(2):73-8.
    PMID: 12804703
    Verification of tumor dose for patients undergoing external beam radiotherapy is an important part of quality assurance programs in radiation oncology. Among the various methods available, entrance dose in vivo is one reliable method used to verify the tumor dose delivered to a patient. In this work, entrance dose measurements using LiF:Mg;Ti and LiF:Mg;Cu;P thermoluminescent dosimeters (TLDs) without buildup cap was carried out. The TLDs were calibrated at the surface of a water equivalent phantom against the maximum dose, using 6- and 10-MV photon and 9-MeV electron beams. The calibration geometry was such that the TLDs were placed on the surface of the "solid-water" phantom and a calibrated ionization chamber was positioned inside the phantom at calibration depth. The calibrated TLDs were then utilized to measure the entrance dose during the treatment of actual patients. Measurements were also carried out in the same phantom simultaneously to check the stability of the system. The dose measured in the phantom using the TLDs calibrated for entrance dose to 6-and 10-MV photon beams was found to be close to the dose determined by the treatment planning system (TPS) with discrepancies of not more than 4.1% (mean 1.3%). Consequently, the measured entrance dose during dose delivery to the actual patients with a prescribed geometry was found to be compatible with a maximum discrepancy of 5.7% (mean 2.2%) when comparison was made with the dose determined by the TPS. Likewise, the measured entrance dose for electron beams in the phantom and in actual patients using the calibrated TLDs were also found to be close, with maximum discrepancies of 3.2% (mean 2.0%) and 4.8% (mean 2.3%), respectively. Careful implementation of this technique provides vital information with an ability to confidently accept treatment algorithms derived by the TPS or to re-evaluate the parameters when necessary.
    Matched MeSH terms: Radiometry/instrumentation*; Radiometry/methods*
Filters
Contact Us

Please provide feedback to Administrator (afdal@afpm.org.my)

External Links