Displaying all 6 publications

Abstract:
Sort:
  1. Reynolds D, Duray GZ, Omar R, Soejima K, Neuzil P, Zhang S, et al.
    N Engl J Med, 2016 Feb 11;374(6):533-41.
    PMID: 26551877 DOI: 10.1056/NEJMoa1511643
    BACKGROUND: A leadless intracardiac transcatheter pacing system has been designed to avoid the need for a pacemaker pocket and transvenous lead.
    METHODS: In a prospective multicenter study without controls, a transcatheter pacemaker was implanted in patients who had guideline-based indications for ventricular pacing. The analysis of the primary end points began when 300 patients reached 6 months of follow-up. The primary safety end point was freedom from system-related or procedure-related major complications. The primary efficacy end point was the percentage of patients with low and stable pacing capture thresholds at 6 months (≤2.0 V at a pulse width of 0.24 msec and an increase of ≤1.5 V from the time of implantation). The safety and efficacy end points were evaluated against performance goals (based on historical data) of 83% and 80%, respectively. We also performed a post hoc analysis in which the rates of major complications were compared with those in a control cohort of 2667 patients with transvenous pacemakers from six previously published studies.
    RESULTS: The device was successfully implanted in 719 of 725 patients (99.2%). The Kaplan-Meier estimate of the rate of the primary safety end point was 96.0% (95% confidence interval [CI], 93.9 to 97.3; P<0.001 for the comparison with the safety performance goal of 83%); there were 28 major complications in 25 of 725 patients, and no dislodgements. The rate of the primary efficacy end point was 98.3% (95% CI, 96.1 to 99.5; P<0.001 for the comparison with the efficacy performance goal of 80%) among 292 of 297 patients with paired 6-month data. Although there were 28 major complications in 25 patients, patients with transcatheter pacemakers had significantly fewer major complications than did the control patients (hazard ratio, 0.49; 95% CI, 0.33 to 0.75; P=0.001).
    CONCLUSIONS: In this historical comparison study, the transcatheter pacemaker met the prespecified safety and efficacy goals; it had a safety profile similar to that of a transvenous system while providing low and stable pacing thresholds. (Funded by Medtronic; Micra Transcatheter Pacing Study ClinicalTrials.gov number, NCT02004873.).
    Matched MeSH terms: Arrhythmias, Cardiac/therapy*
  2. Ritter P, Duray GZ, Zhang S, Narasimhan C, Soejima K, Omar R, et al.
    Europace, 2015 May;17(5):807-13.
    PMID: 25855677 DOI: 10.1093/europace/euv026
    Recent advances in miniaturization technologies and battery chemistries have made it possible to develop a pacemaker small enough to implant within the heart while still aiming to provide similar battery longevity to conventional pacemakers. The Micra Transcatheter Pacing System is a miniaturized single-chamber pacemaker system that is delivered via catheter through the femoral vein. The pacemaker is implanted directly inside the right ventricle of the heart, eliminating the need for a device pocket and insertion of a pacing lead, thereby potentially avoiding some of the complications associated with traditional pacing systems.
    Matched MeSH terms: Arrhythmias, Cardiac/therapy*
  3. Duray GZ, Ritter P, El-Chami M, Narasimhan C, Omar R, Tolosana JM, et al.
    Heart Rhythm, 2017 05;14(5):702-709.
    PMID: 28192207 DOI: 10.1016/j.hrthm.2017.01.035
    BACKGROUND: Early performance of the Micra transcatheter pacemaker from the global clinical trial reported a 99.2% implant success rate, low and stable pacing capture thresholds, and a low (4.0%) rate of major complications up to 6 months.

    OBJECTIVE: The purpose of this report was to describe the prespecified long-term safety objective of Micra at 12 months and electrical performance through 24 months.

    METHODS: The Micra Transcatheter Pacing Study was a prospective single-arm study designed to assess the safety and efficacy of the Micra VVIR leadless/intracardiac pacemaker. Enrolled patients met class I or II guideline recommendations for de novo ventricular pacing. The long-term safety objective was freedom from a system- or procedure-related major complication at 12 months. A predefined historical control group of 2667 patients with transvenous pacemakers was used to compare major complication rates.

    RESULTS: The long-term safety objective was achieved with a freedom from major complication rate of 96.0% at 12 months (95% confidence interval 94.2%-97.2%; P < .0001 vs performance goal). The risk of major complications for patients with Micra (N = 726) was 48% lower than that for patients with transvenous systems through 12 months postimplant (hazard ratio 0.52; 95% confidence interval 0.35-0.77; P = .001). Across subgroups of age, sex, and comorbidities, Micra reduced the risk of major complications compared to transvenous systems. Electrical performance was excellent through 24 months, with a projected battery longevity of 12.1 years.

    CONCLUSION: Long-term performance of the Micra transcatheter pacemaker remains consistent with previously reported data. Few patients experienced major complications through 12 months of follow-up, and all patient subgroups benefited as compared to transvenous pacemaker historical control group.

    Matched MeSH terms: Arrhythmias, Cardiac/therapy*
  4. Afzal MR, Daoud EG, Cunnane R, Mulpuru SK, Koay A, Hussain A, et al.
    Heart Rhythm, 2018 06;15(6):841-846.
    PMID: 29427820 DOI: 10.1016/j.hrthm.2018.02.008
    BACKGROUND: Experience with retrieval of the Micra transcatheter pacing system (TPS) is limited because of its relatively newer technology. Although abandonment of the TPS at end of life is recommended, certain situations such as endovascular infection or device embolization warrant retrieval.

    OBJECTIVE: The purpose of this study was to report the worldwide experience with successful retrieval of the Micra TPS.

    METHODS: A list of all successful retrievals of the currently available leadless pacemakers (LPs) was obtained from the manufacturer of Micra TPS. Pertinent details of retrieval, such as indication, days postimplantation, equipment used, complications, and postretrieval management, were obtained from the database collected by the manufacturer. Other procedural details were obtained directly from the operators at each participating site.

    RESULTS: Data from the manufacturer consisted of 40 successful retrievals of the Micra TPS. Operators for 29 retrievals (73%) provided the consent and procedural details. Of the 29 retrievals, 11 patients underwent retrieval during the initial procedure (immediate retrieval); the other 18 patients underwent retrieval during a separate procedure (delayed retrieval). Median duration before delayed retrieval was 46 days (range 1-95 days). The most common reason for immediate retrieval was elevated pacing threshold after tether removal. The most common reasons for delayed retrieval included elevated pacing threshold at follow-up, endovascular infection, and need for transvenous device. Mean procedure duration was 63.11 ± 56 minutes. All retrievals involved snaring via a Micra TPS delivery catheter or steerable sheath. No serious complications occurred during the reported retrievals.

    CONCLUSION: Early retrieval of the Micra TPS is feasible and safe.

    Matched MeSH terms: Arrhythmias, Cardiac/therapy*
  5. Lloyd M, Reynolds D, Sheldon T, Stromberg K, Hudnall JH, Demmer WM, et al.
    Heart Rhythm, 2017 02;14(2):200-205.
    PMID: 27871854 DOI: 10.1016/j.hrthm.2016.11.016
    BACKGROUND: The Micra transcatheter pacemaker was designed to have similar functionality to conventional transvenous VVIR pacing systems. It provides rate adaptive pacing using a programmable 3-axis accelerometer designed to detect patient activity in the presence of cardiac motion.

    OBJECTIVE: The purpose of this study was to evaluate the system's performance during treadmill tests to maximum exertion in a subset of patients within the Micra Transcatheter Pacing Study.

    METHODS: Patients underwent treadmill testing at 3 or 6 months postimplant with algorithm programming at physician discretion. Normalized sensor rate (SenR) relative to the programmed upper sensor rate was modeled as a function of normalized workload in metabolic equivalents (METS) relative to maximum METS achieved during the test. A normalized METS and SenR were determined at the end of each 1-minute treadmill stage. The proportionality of SenR to workload was evaluated by comparing the slope from this relationship to the prospectively defined tolerance margin (0.65-1.35).

    RESULTS: A total of 69 treadmill tests were attempted by 42 patients at 3 and 6 months postimplant. Thirty tests from 20 patients who completed ≥4 stages with an average slope of 0.86 (90% confidence interval 0.77-0.96) confirmed proportionality to workload. On an individual test basis, 25 of 30 point estimates (83.3%) had a normalized slope within the defined tolerance range (range 0.46-1.08).

    CONCLUSION: Accelerometer-based rate adaptive pacing was proportional to workload, thus confirming rate adaptive pacing commensurate to workload is achievable with an entirely intracardiac pacing system.

    Matched MeSH terms: Arrhythmias, Cardiac/therapy
  6. Coromilas EJ, Kochav S, Goldenthal I, Biviano A, Garan H, Goldbarg S, et al.
    Circ Arrhythm Electrophysiol, 2021 03;14(3):e009458.
    PMID: 33554620 DOI: 10.1161/CIRCEP.120.009458
    [Figure: see text].
    Matched MeSH terms: Arrhythmias, Cardiac/therapy