Affiliations 

  • 1 Universiti Kebangsaan Malaysia
  • 2 Malaysian Institute for Nuclear Technology Research (MINT)
MyJurnal

Abstract

Absorbed dose to water was measured with ionisation chambers NE 2561 (#267), NE 2581 (#334), NE 2571 (#1028), using the IAEA standard water phantom. The ionisation chamber was inserted in the water phantom at a reference depth dependent on the type of the radiation quality used. Three radiation qualities were used namely 1.25 MeV gamma ray, 6 MV x-rays and 10 MV x-rays. The values of the absorbed dose to water were determined by the NK- and NX- based methods, i.e with the use of IAEA, HPA, NACP, AAPM, NCRP and ICRU protocols. The aim of this study was to make an intercomparison of the results, by taking the IAEA protocol as a standard. The largest deviation contributed by any of these protocols was recorded for each quality. It was found that AAPM, NCRP and ICRU protocols contributed 0.94% for 1.25 MeV gamma ray, NACP contributed 2.12% for the 6 MV x-rays, and NACP contributed 2.35% for 10 MV x-rays. Since the acceptable limit of deviation set by the IAEA for this absorbed dose work is ± 3%, it is clear that the overall deviations obtained were all satisfactory.

Similar publications