The paper relates a study on the development of an analysis procedure for measuring the gold coating thickness using EDXRF technique. Gold coating thickness was measured by relating the counts under the Au Lα peak its thickness value. In order to get a reasonably accurate result, a calibration graph was plotted using five gold-coated reference standards of different thicknesses. The calibration graph shows a straight line for thin coating measurement until 0.9μm. Beyond this the relationship was not linear and this may be resulted from the selfabsorption effect. Quantitative analysis was also performed on two different samples of goldcoated jewelry and a phone connector. Result from the phone connector analysis seems to agree with the manufacturer’s gold coating value. From the analysis of gold-coated jewelry it had been able to differentiate the two articles as gold wash and gold electroplated.