Present study concerns the key thermoluminescence (TL) properties of photonic crystal fibres (PCFs), seeking development of alternatively structured TL materials that are able to offer a advantages over existing passive dosimeters. In terms of their internal structure and light guiding properties the PCFs, collapsed and structured, differ significantly from that of conventional optical fibres. To investigate the dosimetric parameters of the PCFs use was made of a linear accelerator producing a 6 MV photon beam, delivering doses ranging from 0.5 Gy to 8 Gy. The parameters studied included TL response, linearity index, glow curves, relative sensitivity and TL signal fading, the results being compared against those obtained using TLD-100 chips. At 4 Gy photon dose the Ge-doped collapsed PCFs were found to provide a response 27 × that of structured PCF, also giving a TL yield similar to that of standard TLD-100 chips. Over post-irradiation periods of 15 and 30 days collapsed PCF TL signal fading were 8% and 17% respectively, with corresponding values of 37% and 64% for the structured PCF. Trapping parameters including the order of kinetics (b), activation energy (E) and frequency factor (s-1) were assessed with Chen's peak shape method. Lifetime of trapping centre was found to be (2.36 E+03) s and (9.03 E +01) s regarding the collapsed and structured PCF respectively with 6 Gy of photon beam. For the Ge-doped collapsed PCF, the high TL yield, sensitivity and low fading provide the basis of a highly promising system of TLD for radiotherapy applications.
* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.