Affiliations 

  • 1 School of Nursing and Midwifery Amol, Mazandaran University of Medical Sciences, Sari, Iran
  • 2 School of Biosciences, Faculty of Health and Medical Sciences, Taylor's University, Subang Jaya, Malaysia
  • 3 Yong Loo Lin School of Medicine, Singapore City, Singapore
  • 4 Kirsehir Ahi Evran University, Faculty of Medicine, Department of Biostatistics and Medical Information Kirsehir, Kirsehir, Turkey
  • 5 Imam Khomeini' s Hospital of Amol, Mazandaran University of Medical Science, Sari, Iran
  • 6 Department of Physiological Nursing, School of Nursing and Department of Epidemiology & Biostatistics, School of Medicine, University of California San Francisco, San Francisco, California, USA
  • 7 Taylor's Business School, Taylor's University Lakeside Campus, Subang Jaya, Malaysia
  • 8 Department of Nursing, Alborz University of Medical Sciences, Karaj, Iran
J Clin Nurs, 2021 Dec 15.
PMID: 34908206 DOI: 10.1111/jocn.16166

Abstract

AIM: To investigate the accuracy, reliability and agreement between infrared forehead thermometers versus infrared tympanic thermometers temperature, a cross-sectional study was conducted in April 2020.

METHODS: The forehead and tympanic temperatures of 615 subjects were measured simultaneously in three exposed SARS-COV-2 groups at one hospital in Iran, during April 2020. These comparisons were evaluated by Bland-Altman Plot, repeatability, Passing-Bablok regression and Lin's concordance correlation coefficient. The receiver operating characteristic (ROC) analysis was done to describe the discrimination accuracy of a diagnostic test. The study adhered to STROBE checklist for cross-sectional studies.

RESULTS: A Bland-Altman plot indicated that the limits of agreement between the forehead and tympanic temperature were -0.259 to +0.19°C. Passing-Bablok regression analysis illustrated that the infrared forehead was not linearly related to tympanic temperatures (reference method), with a slope estimate that was significantly different from 1.00. The infrared forehead thermometer showed poor precision and lower accuracy than the tympanic. The forehead temperature readings had 60.0% sensitivity and 44.4% specificity (p > .05) to predict disease.

CONCLUSION: According to the results of study, there is no evidence that the assessment of temperature by infrared forehead thermometer could discriminate between the two groups (positive and negative).

* Title and MeSH Headings from MEDLINE®/PubMed®, a database of the U.S. National Library of Medicine.