STUDY DESIGN: Prospective study.
METHODS: Looped liners with hook fastener and Iceross Dermo Liner with pin/lock system were mechanically tested using a tensile testing machine in terms of system safety. A total of 10 transtibial amputees participated in this study and were asked to use these two different suspension systems. The pistoning was measured between the liner and socket through a photographic method. Three static axial loading conditions were implemented, namely, 30, 60, and 90 N. Furthermore, subjective feedback was obtained.
RESULTS: Tensile test results showed that both systems could safely tolerate the load applied to the prosthesis during ambulation. Clinical evaluation confirmed extremely low pistoning in both systems (i.e. less than 0.4 cm after adding 90 N traction load to the prosthesis). Subjective feedback also showed satisfaction with both systems. However, less traction at the end of the residual limb was reported while looped liner was used.
CONCLUSION: The looped liner with hook fastener is safe and a good alternative for individuals with transtibial amputation as this system could solve some problems with the current systems. Clinical relevance The looped liner and hook fastener were shown to be good alternative suspension for people with lower limb amputation especially those who have difficulty to use and align the pin/lock systems. This system could safely tolerate centrifugal forces applied to the prosthesis during normal and fast walking.
CASE DESCRIPTION AND METHODS: The subject had unstable excessive soft tissue at the distal end of the residuum. After 2 years of prosthetic use, she experienced several problems, including pain in the residual limb and knee joint. Pressure distribution was evaluated during ambulation. We also designed a total surface bearing socket with Velcro as suspension system to distribute the load evenly on the residual limb, and to facilitate donning procedure.
FINDINGS AND OUTCOMES: The main site of weight bearing in the old prosthesis (patellar tendon bearing) was anterior proximal region of the residual limb, especially the patellar tendon. The pressure was almost 10 times higher than the distal region during level walking. Pressures were distributed more evenly with the total surface bearing socket design, and the donning was much easier.
CONCLUSION: Pressure distribution within the socket could be affected by socket design and suspension system. Using the total surface bearing socket and Velcro as suspension system might facilitate donning of prosthesis and reduce traction at the end of residual limb during the swing phase of gait.
CLINICAL RELEVANCE: Proper socket design and suspension system based on the amputees' need can facilitate rehabilitation process and lead to the amputee's satisfaction. The pressure is distributed more uniformly over the residual limb by the total surface bearing socket compared to the patellar tendon bearing socket for lower limb amputees with unusual stump shape.
Patients and methods: This study included 50 unilateral transtibial amputees (26 males, 24 females; mean age 55.4±14.7 years; range, 18 to 78 years) who met the respondent criteria of the study. The respondents were randomly selected and were issued with a questionnaire.
Results: The majority of the transtibial amputees were aware of stump contracture complications. It was found that they also preferred methods of prevention which required less effort, was cost-effective, and were also practical.
Conclusion: More focus should be placed on contracture prevention methods which were most complied with by the patients. This is because the effectiveness of a prevention method is highly influenced by patients' compliance with the method. Patients with a higher risk of developing stump contractures should be also given more attention in post-amputation care.